MRO2X
– 5X Manual
- 38 -
98-0058
Rev. R
8.2
PT401 ANTI-SCALANT
Smaller AmeriWater MRO products have the PT401 antiscalant system equipped as an option.
The larger units, MRO2-5X4 ordinarily have a complete water pre-treatment system
(backwashing particulate filter, backwashing carbon filter, and water softener) installed ahead of
the MRO. This eliminates the need for the PT401 anti-scalant system within the MRO, and is
not a point of discussion in this manual. If your unit has been fitted with a PT401 anti-scalant
system due to a water pre-treatment system not being installed ahead of the MRO, refer to the
PT401 manual.
8.3
MEMBRANE MAINTENANCE INSTRUCTIONS
MEMBRANE REPLACEMENT PROCEDURE:
Turn off the incoming tap water supply to the MRO and unplug the device from the GFI
receptacle. Power down the disconnect for the motor starter. (The following procedure will
cause water to leak and spill
– This could occur at different times during the procedure - You
may get wet.)
1.
Place a container under the CHLORAMINE SAMPLE (TEST) PORT and slowly open the
port to relieve the pressure on the MRO system.
2.
Place a container under the PRODUCT SAMPLE (TEST) PORT on the front panel and
open to allow to drain.
3.
Open the back panel of cabinet.
4.
Press release button on Product Water Manifold (Manifold located near top at back of left
wall
– looking in back of cabinet) to disconnect product water hose from all membranes
being removed.
5.
Disconnect the pump feed and reject discharge hose from the membrane being
replaced.
6.
Remove the unistrut clamps from membrane and pull the membrane assembly from the
cabinet.
7.
With membrane removed, loosen inlet cap clamp and remove inlet cap from the
membrane assembly.
8.
Remove used membranes from the housing.
9.
Note the serial number of the replacement membrane and indicate this on the outer
housing of the membrane assembly with a label maker.
10.
Insert the new membrane into the housing inlet.
NOTE:
You must insert membranes into the inlet end of the housing. Inserting a
membrane from the discharge end will damage the membrane’s brine seal.
Summary of Contents for MRO2X
Page 2: ......
Page 9: ...MRO2X 5X Manual 5 98 0058 Rev R SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW ...
Page 11: ...MRO2X 5X Manual 7 98 0058 Rev R 3 2 INTERNAL REAR VIEW ...
Page 13: ...MRO2X 5X Manual 9 98 0058 Rev R 3 3 ELECTRICAL DIAGRAM MRO2X MRO3X MRO4X MRO5X ...
Page 14: ...MRO2X 5X Manual 10 98 0058 Rev R 3 4FLUID DIAGRAM MRO2X MRO3X MRO4X ...
Page 15: ...MRO2X 5X Manual 11 98 0058 Rev R 3 5 FLUID DIAGRAM MRO5X ...
Page 35: ...MRO2X 5X Manual 31 98 0058 Rev R FIGURE 6 2 ...
Page 36: ...MRO2X 5X Manual 32 98 0058 Rev R FIGURE 6 3 ...