MRO2X
– 5X Manual
- 25 -
98-0058
Rev. R
5.4
MEMBRANE FLUSH FEATURE (AUTO FLUSH)
The MEMBRANE FLUSH FEATURE is the preferred means for minimizing bacterial growth for
the MRO2-5X during periods of low water usage. Ordinarily these sizes of MRO are connected
to a loop or storage tank only having 1-2 days of non-use when dialysis procedures are not
being carried out. The MRO2-5X can be set up to discourage microbiological growth by
“flushing” periodically.
The Membrane Flush feature is disabled as a default from the factory for MRO2-5X models.
Refer to the default setpoints in Section 6.4. If these default settings do not meet your
particular need, then they can be changed. See Section 6.5 on how to change from the default
settings.
FLUSH MODE
RO PUMP
INLET VALVE
DIVERT VALVE
3
(FOR STORAGE TANK APPLICATIONS)
ON
OPEN
ENABLED
4
(FOR LOOP APPLICATIONS)
ON
OPEN
DISABLED
FLUSH MODE #3
assumes that the
MRO is connected to a storage tank
and, therefore, will
divert the PRODUCT WATER to the drain. This minimizes bacterial growth and directs all the
water to the drain, rather than refill the storage tank.
FLUSH MODE #3
will cause all the water
to be diverted to drain for each MEMBRANE AUTO FLUSH cycle.
On the other hand, when an
MRO is connected to a Direct Feed Loop
, the PRODUCT water
should be directed into the loop during the MEMBRANE AUTO FLUSH cycle to keep the water
in the loop “fresh”. Setting the
FLUSH MODE for #4
will disable the DIVERT TO DRAIN
feature, and the PRODUCT WATER will be circulated through the loop during each
MEMBRANE AUTO FLUSH cycle. The MEMBRANE AUTO FLUSH will operate while the MRO
is in the STANDBY mode, while connected to the loop even though no water is being used for
dialysis.
SETPOINT SETTING
Flush Type
6 (Off hours)
Flush Time
15 to 45 (minutes)
Flush Mode
3 (Inlet valve open, pump on)
Flush Interval
4 to 12 (hours)
The MRO must be connected to the electric power source, incoming water supply, and drain at
all times for this feature to be operable.
The Flush Time and Flush Interval settings are recommended settings, but may be adjusted to
fit your specific needs. Contact your AmeriWater representative for guidance.
Summary of Contents for MRO2X
Page 2: ......
Page 9: ...MRO2X 5X Manual 5 98 0058 Rev R SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW ...
Page 11: ...MRO2X 5X Manual 7 98 0058 Rev R 3 2 INTERNAL REAR VIEW ...
Page 13: ...MRO2X 5X Manual 9 98 0058 Rev R 3 3 ELECTRICAL DIAGRAM MRO2X MRO3X MRO4X MRO5X ...
Page 14: ...MRO2X 5X Manual 10 98 0058 Rev R 3 4FLUID DIAGRAM MRO2X MRO3X MRO4X ...
Page 15: ...MRO2X 5X Manual 11 98 0058 Rev R 3 5 FLUID DIAGRAM MRO5X ...
Page 35: ...MRO2X 5X Manual 31 98 0058 Rev R FIGURE 6 2 ...
Page 36: ...MRO2X 5X Manual 32 98 0058 Rev R FIGURE 6 3 ...
Page 52: ...MRO2X 5X Manual 48 98 0058 Rev R SECTION 11 SPARE PARTS LISTING ...