8100 LAL
Blower Unit Cleaning Instructions:
WARNING and CAUTION:
A. Hand clean only. DO NOT place in sterilization room or chamber.
B. It is recommended that gloves and protective clothing be worn at all times during cleaning
and disinfecting
C. Hypercarbonate and phenol based solutions
should not
be used.
D. Make sure unit is turned “off” and unplugged before cleaning.
For normal soil, wipe down with a standard cleaning spray and wipe off excess moisture with a towel.
Spray with a disinfectant and let it stand for a few minutes before wiping down. Check corners and
creases for soil build-up. Look to see if any external parts are damaged or need replacing such as:
hooks, foot pads and filter. Replacement parts can be issued. Contact your representative for servicing.
NOTE:
To keep your equipment working efficiently and effectively for an extended time period, it is
essential to make sure the equipment is cleaned regularly and well maintained.
Maintenance:
Air Filter
The foam air filter on the back of the Control Unit must be cleaned
weekly with warm soap and water. Replacement of the foam filter is
recommended every 6 months.
Storage:
Deflate mattress until all air is exhausted completely. With the cover on, roll the mattress starting at the
foot section. Secure with straps and keep mattress in a clean dry area away from heat or flames. Store
unit and mattress in a temperature range between 40°and 95°. DO NOT expose to moisture or areas of
extreme humidity.
Disposition:
Once the unit has reached the end of its useable life, contact your representative for service, disposition
or replacement.
The filter is easily removed and
reinserted through the gap in the back
case.