American National Manufacturing Static LAL 8100 User Instruction Manual Download Page 10

 

10 of 16 

 

8100 LAL 

Mattress Cleaning Instructions: 

 
 
WARNING and CAUTION: 

 

A.  It is recommended that gloves and protective clothing be worn at all times during cleaning 

and disinfecting 

B.  Hypercarbonate and phenol based solutions 

should not

 be used. 

 
During Operation: 
 

For normal soil, wipe down cover and interior sacs with a standard cleaning disinfectant and 
wipe off excess moisture with a towel. 

 
Laundering Mattress: 

1.  Disconnect the power from the Control Unit and deflate the interior sacs. 
2.  Disassemble cover, interior sacs, connecting hoses, and base.  Place them in different 

laundering baskets. 

3.  The interior sacs must be hand cleaned only.  DO NOT place in sterilization room or 

chamber.  Use a spray disinfectant to wipe down interior sacs.  Air-dry sacs. 

DO NOT 

expose to heat.

 

4.  Spray disinfectants and wipe down foam cover. 
5.  Place cover and base

 ONLY

 in washing machine. 

6.  Use standard hospital disinfectant/detergent. 
7.  Set washing cycles as recommended below: 

Regular Cycle

 

a)  Pre-soak with disinfectant/detergent with diluted bleach in cold water (10 

minutes.) 

b)  Main wash for 15 minutes (Time depending on soil level). 
c)  Rinse Cycle: 5 minutes minimum. 
d)  Spin/Drain Cycle: 5 minutes minimum. 
e)  Air-dry or place in dryer at very low or no heat. 

8.  After the cover, interior sacs, and base are dry, store them in a sterilized bag and seal for 

next use. 

 

NOTICE:

  DO NOT use water temperatures greater than 150°F (65°C). 

Summary of Contents for Static LAL 8100

Page 1: ...Static Low Air Loss Mattress Replacement System Read Entire Manual before Operating Device ANSLAL Uncontrolled Document Rev 1 0 08 26 2008 User Instruction Manual...

Page 2: ...arnings and Precautions 4 Seven Zones of Bed Rail Entrapment 5 Device Information 6 Unpacking and Set up 7 Operating Instructions 8 Keypad Quick Reference 9 Mattress Cleaning Instructions 10 Blower Un...

Page 3: ...ient or user injury A CAUTION points out special procedures or precautions that personnel must follow to avoid equipment damage This symbol highlights an ELECTRICAL SHOCK HAZARD WARNING Symbols Found...

Page 4: ...out of direct sunlight CAUTION DO NOT store in temperatures below 40 or greater than 95 Fahrenheit 5 35 Celsius CAUTION DO NOT expose to moisture or areas of extreme humidity CAUTION Ensure that stra...

Page 5: ...em the caregiver is responsible for seeing that the mattress properly fits the bed frame It is also the caregiver s ultimate decision whether or not to use bed rails with the patient Zone One Zone Two...

Page 6: ...l panel that has a turn dial for pressure adjustment Inside the Control Unit is a variable output blower that allows the air mattress to operate in static mode There is also a printed circuit board wh...

Page 7: ...s Fahrenheit DO NOT expose to moisture or areas of extreme humidity Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cable...

Page 8: ...7 Turn master power switch located on the side of the unit on 8 Press the Autofirm button for quick inflation 9 Press the Firm or Soft button on the control panel to select the desired level of firmn...

Page 9: ...n Lockout Locks all functions auto matically after 3 mins To disable press and hold lockout button for 3 sec Firm Increases airflow for a firmer setting Alarm Silence Mutes the audible alarm Visual li...

Page 10: ...interior sacs must be hand cleaned only DO NOT place in sterilization room or chamber Use a spray disinfectant to wipe down interior sacs Air dry sacs DO NOT expose to heat 4 Spray disinfectants and w...

Page 11: ...our representative for servicing NOTE To keep your equipment working efficiently and effectively for an extended time period it is essential to make sure the equipment is cleaned regularly and well ma...

Page 12: ...of the hoses are punctured kinked or disconnected Check for proper connections from the hoses to the pump Make sure they are secure Check or clean intake filters Be sure CPR valve on the mattress is c...

Page 13: ...unit to be hung on bed frame Female CPC Connects from mattress to Male CPC on blower unit Removes exhaust air from unit to mattress Male CPC Chrome Attached to blower unit side panel Connects to Femal...

Page 14: ...ons Warranty American National Manufacturing Inc Warrants that all products provided under this agreement are free from defects in material and workmanship for the following stated warranty period fro...

Page 15: ...15 of 16 8100 LAL Notes...

Page 16: ...m All specifications equipment and prices are subject to change without notice Photos and drawings are representative of the products and may vary slightly from actual production models Contact or con...

Reviews: