Exhaust level
The exhaust level can be changed from level 0 to level 3 in the liquid and agar modes.
It can be changed also during operation.
①
Press the ► switch to increase the exhaust level, and press the ◄ switch to decrease
the level.
EXHT.
●
Rapid exhaust after sterilization of liquid may cause the liquid to spill. Set a lower exhaust level
for fine exhaust, or set the level 0 (natural cooling).
●
Continuous operation at the exhaust level of 3 may cause steam to come out of the exhaust port
of the drain bottle. In that case, lower the exhaust level, except for the solid cycle, and turn ON
the fan cooling function.
●
Fan cooling ON/OFF
Except the solid mode, fan cooling ON/OFF can be changed. It can be changed also during
operation.
①
Press the "Fan cooling ON/OFF" switch.
・
The fan rotates after sterilization when the fan indicator is lighting.
FAN
ON
OFF
△
NOTE