B. RX 

Mode 

 
Without a Chest Piece Assembly plugged into the Chest Piece jack of the TR-1, the unit 
is automatically in the Receive Mode.  With the Data Communications Channel 
established and a Headset plugged into the Headset jack of the unit, the clinician is set 
up to hear the stethoscope sounds from the remote AMD-3700 (in Transmit Mode). 
 
Volume to the Headset can be increased by turning the Volume control clockwise.  
Increasing the volume too high can cause distortion in the Headset and will also 
increase the background noise.  Start with the lowest volume necessary to hear the 
sounds, then turn up the volume as needed to focus in on specific sounds. 
 
The B/D Switch can be used in an auscultation exam similarly to a traditional 
stethoscope.  The Bell (B) position cuts off the high end at 250 Hz and boosts the low 
frequency sounds below 100 Hz for enhanced heart auscultation.  The Diaphragm (D) 
position slightly reduces the very low frequencies and extends the high frequencies to 
1,400 Hz. 
 

IV. 

Maintenance and Calibration 

 
There is no scheduled maintenance required for the AMD-3700 Telephonic 
Stethoscope. 
 
There is no calibration required for the AMD-3700 Telephonic Stethoscope. 
 
If problems should arise with the AMD-3700 Telephonic Stethoscope, the table in the 
following section may be used to trouble shoot the problem.  Any failed units should be 
returned for repair. 
 
The AMD-3700 Telephonic Stethoscope may be cleaned with a damp cloth. 
 

AMD Global Telemedicine, Inc. 

73 Princeton Street   North Chelmsford, MA 01863 

978-937-9021 (phone)   978-937-5249 (fax) 

6

Summary of Contents for AMD-3700

Page 1: ...D Global Telemedicine Inc 73 Princeton Street North Chelmsford MA 01863 978 937 9021 phone 978 937 5249 fax AMD Global Telemedicine Inc Installation and Operation Manual AMD 3700 Telephonic Stethoscop...

Page 2: ...flammable anesthetic mixture with air or with continuous oxygen or nitrous oxide This product complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device...

Page 3: ...lection The Bell B position cuts off the high end at 250 Hz and boosts the low frequencies below 100 Hz for enhanced heart auscultation The Diaphragm D position slightly reduces the very low frequency...

Page 4: ...B 9 straight through cable should be used Sometimes DCE equipment will have a DB 25 connector for its interface In that case a DB 9 to DB 25 adapter may be used or a cable with a female DB 9 connector...

Page 5: ...signals and sent out over the RS232 data interface In addition to transmitting the digital signals the auscultation sounds from the Chest Piece are looped back and presented at the local Headset jack...

Page 6: ...used in an auscultation exam similarly to a traditional stethoscope The Bell B position cuts off the high end at 250 Hz and boosts the low frequency sounds below 100 Hz for enhanced heart auscultation...

Page 7: ...placement The Chest Piece is plugged in the TX Mode light is lit but there is no sound heard in the local Headset The Chest Piece Assembly is not plugged in all the way Make sure Chest Piece Assembly...

Reviews: