background image

If sounds from the Chest Piece are heard at the headset, then the AMD-3700 units are 
working properly and if problems arise in connecting the unit to the Data Channel, it is 
appropriate to the concentrate on getting the signal through the Data Communications 
Channel. 
 

C. Mode 

Selection 

 
If there is no Chest Piece Assembly plugged into the Chest Piece jack of the unit, then 
the unit will be in Receive Mode and the RX Mode light will be illuminated.  The TX 
Mode light will be OFF. 
 
If a Chest Piece Assembly is plugged into the Chest Piece jack of the unit, then the unit 
will be in Transmit Mode and the TX Mode light will be illuminated.  The RX Mode light 
will be OFF. 
 

III. Operation 

 
For a normal remote auscultation exam, the AMD-3700 unit located with the patient is in 
Transmit Mode and the AMD-3700 unit with the clinician is in the Receive Mode. 
 

A. TX 

Mode 

 
To operate as an auscultation transmitting unit the AMD-3700 must have a Chest Piece 
Assembly plugged into the Chest Piece jack.  That will put the unit into Transmit Mode 
and illuminate the TX Mode light located just over the Chest Piece jack.  Sounds from 
the Chest Piece will be converted to digital signals and sent out over the RS232 data 
interface. 
 
In addition to transmitting the digital signals, the auscultation sounds from the Chest 
Piece are looped back and presented at the local Headset jack.  Thus, with an optional 
Headset, the local user can listen to the same auscultation sounds that the listener at 
the remote AMD-3700 (in Receive Mode) would hear.   
 
Volume to the Headset can be increased by turning the Volume control clockwise.  
Increasing the volume too high can cause distortion in the Headset and will also 
increase the background noise.  Start with the lowest volume necessary to hear the 
sounds, then turn up the volume as needed to focus in on specific sounds. 
 
The B/D Switch can be used in an auscultation exam similarly to a traditional 
stethoscope.  The Bell (B) position cuts off the high end at 250 Hz and boosts the low 
frequencies below 100 Hz for enhanced heart auscultation.  The Diaphragm (D) position 
slightly reduces the very low frequency sounds and extends the high frequencies to 
1,400 Hz. 
 
Adjusting the Volume or changing the B/D Switch only affects what the local listener 
hears and does not affect what the remote AMD-3700 listener would hear.  The remote 
AMD-3700 listener has independent Volume and B/D Switch controls. 
 

AMD Global Telemedicine, Inc. 

73 Princeton Street   North Chelmsford, MA 01863 

978-937-9021 (phone)   978-937-5249 (fax) 

5

Summary of Contents for AMD-3700

Page 1: ...D Global Telemedicine Inc 73 Princeton Street North Chelmsford MA 01863 978 937 9021 phone 978 937 5249 fax AMD Global Telemedicine Inc Installation and Operation Manual AMD 3700 Telephonic Stethoscop...

Page 2: ...flammable anesthetic mixture with air or with continuous oxygen or nitrous oxide This product complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device...

Page 3: ...lection The Bell B position cuts off the high end at 250 Hz and boosts the low frequencies below 100 Hz for enhanced heart auscultation The Diaphragm D position slightly reduces the very low frequency...

Page 4: ...B 9 straight through cable should be used Sometimes DCE equipment will have a DB 25 connector for its interface In that case a DB 9 to DB 25 adapter may be used or a cable with a female DB 9 connector...

Page 5: ...signals and sent out over the RS232 data interface In addition to transmitting the digital signals the auscultation sounds from the Chest Piece are looped back and presented at the local Headset jack...

Page 6: ...used in an auscultation exam similarly to a traditional stethoscope The Bell B position cuts off the high end at 250 Hz and boosts the low frequency sounds below 100 Hz for enhanced heart auscultation...

Page 7: ...placement The Chest Piece is plugged in the TX Mode light is lit but there is no sound heard in the local Headset The Chest Piece Assembly is not plugged in all the way Make sure Chest Piece Assembly...