Table 5. EN 60601-1-2:2015 Electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
Required Test
60601 test level for
equipment used in a
professional healthcare
facility only
Compliance level
Surge
IEC61000-4-5
± 0.5, 1, 2 kV line(s) to earth,
± 0.5, 1 kV line(s) to line(s) for
power supply lines
± 2 kV line(s) to earth for input /
output lines
± 0.5, 1, 2 kV line(s) to earth,
± 0.5, 1 kV line(s) to line(s) for
power supply lines
Conducted RF
IEC61000-4-6
3 V - 150 kHz to 80 MHz
6V - ISM bands between 150
kHz to 80 MHz
3 V - 150 kHz to 80 MHz
6V - ISM bands between 150
kHz to 80 MHz
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
0% UT
(100 % dip in UT)
For 0.5 cycle
0% UT
(100 % dip in UT)
For 1 cycle
70 % UT
(30 % dip in UT)
for 25/30 cycles
0% UT
(100 % dip in UT)
For 250/300 cycles
0% UT
(100 % dip in UT)
For 0.5 cycle
0% UT
(100 % dip in UT)
For 1 cycle
70 % UT
(30 % dip in UT)
for 25/30 cycles
0% UT
(100 % dip in UT)
For 250/300 cycles
Power frequency (50/60Hz)
Magnetic field
IEC61000-4-8
30 A/m
30 A/m