ALT I076 Issue 03/0418
Page 15
Part
Description
Item
Numbers*
Checks to be carried out
Bottle
4
Ensure there is no physical damage to the bottle, cracks, splits,
holes etc., and that the threads are not damaged and that the
structure is not deformed. Dispose of any failed bottles and
replace with new ones.
4
Ensure the bottle is clean and dry and does not have residues in
the bottom. If dirty clean, may be sterilized if deemed necessary.
Power
Supply and
Lead
10 and 11
Check the grommets, connectors, plugs and wire for any signs of
physical damage, e.g. exposed wires, splits in the cable, cracks in
casing etc. Do not use if there is any sign of damaged
10 and 11
Make sure the connection between the lead and power supply
unit is secure, if loose push together to ensure the connection is
secure.
Q-Rinse
Machine
N/A
Check for any physical damage to the casing e.g. cracks, chips
etc., or other physical damage to the valves or switches or
connectors. If damaged replace.
N/A
Ensure the four rubber feet are in place and secure to prevent
movement of the machine. If missing replace.
15
Check to make sure the 10 adaptors are present and ready for
use if needed. If missing replace.
Assembled
Q-Rinse
System
N/A
Ensure the waste receptacle is designed to accept contaminated
fluid waste. If not designated for contaminated waste, source a
suitable receptacle approved by your facility.
6 and 16
Ensure the blue tubing goes into the Fluid Valve and the clear
tubing goes into the Air Valve. Ensure the elbow connector on
the green tubing has clicked securely into place. If not set up
accordingly and re-read the Instruction Manual.
10 and 11
Ensure the Power Supply and lead cannot be splashed or
exposed to water. Use splash protection over sockets if present.
N/A
Run an automatic cycle to ensure fluid and air flow is correct and
that the switch over between fluid and air takes place. Do not
use if the cycle fails, send for repair.