Auto-Pure Mini Operation Manual

Chapter 4 Operations

25

rod action

Cooling fan

Optional on or off, off by default

Dry

settings

Above the kit

When the wait step is

performed, the bottom of the

magnetic rod's tip is placed

above the kit

The default drying

position is on the top

of the kit;The default

is 5mm above the

liquid level;

Specified location

When the wait step is

performed, the bottom of the

magnetic rod's tip is located at

the set height above the liquid

level

Action

parameter

Horizontal velocity

(1-50mm)

Set the left/right movement speed, default 50

Lifting speed of

magnetic

rod

(1-100mm)

Set up and down speed of magnetic rod, default 50

Lifting speed of

magnetic rod's tip

(1-100mm)

Set up and down speed of magnetic rod's tip, default

50

6.1.5

Quick new protocol

Click "New" button on the home page to quickly create a new protocol (the

method of creating a new protocol is the same as 6.1.4)

Fig 32 Quick new protocol

Summary of Contents for Auto-Pure Mini

Page 1: ...Operation Manual V1 0 Auto Pure Mini Nucleic Acid Purification System Hangzhou Allsheng Instruments Co Ltd...

Page 2: ...iculties Opening Check Please check the Instrument and Accessories according to the packing list when you first open the packing case If anything wrong or missing please contact the distributor or the...

Page 3: ...is normal indoor instrument which conforms to class of GB 4793 1 standard Please read this manual carefully before operation The device must be used by experienced personnel with appropriate training...

Page 4: ...iod of time cover it with a cloth or plastic to protect it from dust Disconnect the power cord from the jack at once in the following cases and contact your local distributor or Allsheng Liquid enters...

Page 5: ...ive The products under warranty must be delivered by customer to the repair department designated by our company The freight of the instrument from the user to the maintenance department shall be paid...

Page 6: ...on 3 Chapter 3 Basic Operating Instructions 4 1 Structure 4 2 Operation Panel 6 Chapter 4 Operations 7 1 Preparation before use 7 2 Power Connection 8 3 Kits Installation 8 4 Magnetic rod s tip instal...

Page 7: ...Contraindication No contraindication 3 Service Life Service life of the instrument is five years For production date please see the label on back of the instrument Features Humanized operation English...

Page 8: ...ction reagents 10 60min time Temperature control module Ambient temperature 120 for lysis and elution Heating time Ambient temperature 120 5 minutes Temp Accuracy 1 Operation 4 3 inch touch screen APP...

Page 9: ...plate special single sample kit Purification disinfection Fan exhaust UV disinfection Power Supply 100 240V 10 50 60Hz 5 support battery power Power Standby 10W MAX 150W Operating temperature 10 40 Op...

Page 10: ...Chapter 3 Basic Operating Instructions This chapter mainly introduces structures basic operation keys displays as well as preparations before starting up Please read this chapter carefully before usi...

Page 11: ...Auto Pure Mini Operation Manual Chapter 3 Basic Operating Instructions 5 Fig 3 Back 1 3 Side Fig 4 Side Fan Power Switch RS232 Port Power Socket Fuse USD Port Fuse PC Port...

Page 12: ...screen mouse also can be connected for operation TAB Switching between protocols RUN Start the protocol and run the instrument STOP Stop the operation STOP Lighting UV lamp top of test area State lamp...

Page 13: ...packing case Remove the tape holding the cabin door open the door remove the fastening belt between the magnetic rod holder and the Magnetic rod s tip holder Then unscrew the hexagonal screw fixing th...

Page 14: ...g strip according to the arrow and place it smoothly Note the orientation of the plate kit corner A1 corresponds to the marker A1 on the instrument to avoid the wrong sequence of reagents After the pl...

Page 15: ...s tip installation Insert the magnetic rod s tip into the fixed groove of the magnetic rod s tip holder in the direction of the arrow to the end The number of magnetic rod s tip can be determined acco...

Page 16: ...only supported by Android devices used to complete the edit of experimental protocols generation of QR code and other operations The instrument software is oriented to the experimenter and is used for...

Page 17: ...face After the startup is completed you will enter the startup interface where the software version information will be displayed Then into the main interface whose upper part display the name and inf...

Page 18: ...g interface Internet of Things Click into Internet function interface 6 1 3 Protocol interface Click protocol on main interface to enter the protocol interface protocol interface is mainly divided int...

Page 19: ...Operation Manual Chapter 4 Operations 13 Fig 15 Protocol list interface The protocol interface mainly displays the protocol list with new preview selection protocol search and other function Fig 16 Pr...

Page 20: ...four buttons appear at the bottom of the interface Transmission Delete Protocol code and Save as which can carry out Bluetooth transmission protocol Delete generate the QR code of the protocol and sa...

Page 21: ...tions 15 Fig 18 1 protocol selected Fig 19 Multiple protocol selected When multiple protocols are selected only Bluetooth transmission and protocol deletion can be performed Transmission Transfer the...

Page 22: ...Protocol transmission Fig 21 Protocol deletion Delete Removal unnecessary protocol Save as The selected single protocol can be saved as a new protocol default save protocol name is current time seria...

Page 23: ...n Protocol QR code Click QR code Pop up code generation interface every 15 steps to generate a QR code operation under a single protocol The code is stored in an album as shown in Fig 22 Save the form...

Page 24: ...ol Click the New icon on the protocol interface to customize the protocol name The default protocol name is the time For example The current time is 7 14 so the first protocol is 07141 the second prot...

Page 25: ...Auto Pure Mini Operation Manual Chapter 4 Operations 19 Fig 26 New step Fig 27 Custom steps added...

Page 26: ...the parameters of the steps and generate the two dimensional code of the protocol Add step New procedure steps you can use the App s own step name or customize the edit according to personal habits Pr...

Page 27: ...lete and copy 6 1 4 1 Step parameters settings Click a single step to enter the parameter setting interface as shown in Table 4 Only 1 and 6 hole have heating function the top display is the current s...

Page 28: ...Auto Pure Mini Operation Manual Chapter 4 Operations 22 Fig 29 General parameter settings Fig 30 Advanced parameter...

Page 29: ...Temperature Set the temperature of the hole in the range of 37 120 1 and 6 hole have heating function Default OFF Segment The segments of magnet can be set in the range of 1 5 Every time Can be set f...

Page 30: ...out synchronously with the magnetic rod s tip or magnetic rod action Action after reach specified T When running to the heating hole position the magnetic rod s tip or magnetic rod start action after...

Page 31: ...step is performed the bottom of the magnetic rod s tip is located at the set height above the liquid level Action parameter Horizontal velocity 1 50mm Set the left right movement speed default 50 Lif...

Page 32: ...unning log list of the instrument and download them The downloaded protocol can be viewed and edited in the Protocol Instrument Acquisition After downloading the operation log you can click the previe...

Page 33: ...Auto Pure Mini Operation Manual Chapter 4 Operations 27 Fig 34 Acquire the protocol list Fig 35 Protocol download...

Page 34: ...eiving APP are The protocol name is the same The SN number of the instrument is the same if two are met the duplicate name is considered 6 1 6 2 Running log Select the running log that you want to dow...

Page 35: ...Auto Pure Mini Operation Manual Chapter 4 Operations 29 Fig 37 Running log download Fig 38 Log preview...

Page 36: ...e Bluetooth is turned on on the mobile device the APP can directly open the Bluetooth My instrument refers to all the instruments that have been connected by the APP at present The right side will sho...

Page 37: ...2 Modify instrument information Click on instrument interface to enter modify instrument information interface as shown in figure 42 Fig 42 Modify instrument information Click Name to modify the note...

Page 38: ...art up Interface Turn on the instrument start up interface will comes up Fig 43 Start up Interface 6 2 2 Protocol running After self check enter the protocol run interface the interface is divided int...

Page 39: ...l function buttons of run interface are shown in the table below Table 6 Run interface function buttons Icon Function Delete used to delete protocol Shortcut for switching between quick display and li...

Page 40: ...protocol the instrument automatically generates and stores an running report which contains name time step information alarm record and instrument information etc and can be exported The running repor...

Page 41: ...cted report Search used to search for the specified report Export can be exported to U disk 6 2 5 Settings interface The Settings interface mainly has four functions UV Fan settings and Instrument UV...

Page 42: ...Operation Manual Chapter 4 Operations 36 Fig 49 UV sterilization Fan users can choose On or Off Fig 50 Fan Settings interface is divided into Time Language Transfer Bluetooth and Version as shown in t...

Page 43: ...Auto Pure Mini Operation Manual Chapter 4 Operations 37 Fig 51 System Settings Fig 52 Settings Time Settings as shown in the figure below Fig 53 System time...

Page 44: ...Mini Operation Manual Chapter 4 Operations 38 Language Setting support in Chinese and English Fig 54 Language Settings After inserting the U disk import or export the program by click Transfer Fig 55...

Page 45: ...nstrument Fig 56 Bluetooth System maintenance Settings this function is only open to after sales maintenance personnel the instrument has been adjusted before leaving the factory the user does not nee...

Page 46: ...ure Replace light tube Contact with distributor 4 Can not stop automatically after opening the door Sensor failure Contact with distributor 5 Big variance between actual and display temperature Sensor...

Page 47: ...E015 E025 E035 E045 3 T1 T2 T3 T4 short circuit E015 E025 E036 E046 4 Horizontal zero sensor damaged E403 5 Magnetic rod s tip motor zero sensor damaged E425 6 Magnetic rod motor zero sensor damaged E...

Page 48: ...ce only may appear in the above user manual A Ampere AC Alternating Current V Volt Hz Hertz W Watt USB Universal Serial Bus SD Safety Digital Code WiFi Wireless LAN kg Kilogram mm Micrometer L Microli...

Page 49: ...Auto Pure Mini Operation Manual Chapter 6 Acronyms and Marks 43 2 Marks The following marks appear on the device Warning Heating European standard Beware of clamping hand...

Page 50: ...er motor moves left and right Stepper motor magnetic rods tip moves and motor lock Stepper motor magnetic rods moves and motor lock Photoelectric switch Photoelectric switch Photoelectric switch Input...

Page 51: ...Memo...

Reviews: