Auto-Pure Mini Operation Manual
Chapter 4 Operations
23
The parameters of each step are shown in the following table:
Table 4 Step information description
Name
Meaning
Save
Click to save settings
Name
Name of step. A step is not allowed to be renamed. Only a
custom step can be renamed. Custom step parameters are
shown in Fig 27.
Hole site
The hole site is 1-6, the default is 1 for lysis, 2 for collect
beads, 1 for bind, 3 for wash, 6 for elution, and 3 for discard
Volume
Set the volume of the corresponding hole site (5-1000ul)
Mix time
Set the mix time between 0.0 and 99.9min. You can enter a
decimal digit.
Mix speed
Set the mix speed to a range of 1-10.
Dry time
Set dry time, rang 0.0-99.9min.
Temperature
Set the temperature of the hole in the range of 37-120
℃
, 1
and 6 hole have heating function. Default OFF
Segment
The segments of magnet can be set in the range of 1-5.
Every time
Can be set for each period of time, range 0-999s.
Magnetization time
Set magnetization times. Range 0-999s.
Cycle
The number of cycle can be set.The range is 1-10.
Magnet speed
Set magnet speed, range 0.1-15.0mm/s.
Total time
Settings are not supported, calculated
Mix scope
Set mix scope. Range 10-100%, the default is 80%
Mix pos.
Set mix position. Range 0-100%, the default is 0%
Magnet pos.
Set magnet position. Range 0-100%, the default is 0%
Click Parameter Save to save the current step setting parameters.
6.1.4.1 Option settings
Click
at the top right corner of step interface to enter into optional
interface for senior parameters settings. Then click “save” after finishing
settings, as shown in the figure below;