Auto Pure 24/48/96 Operation Manual
Chapter 4 Operations
18
Fig 23
Segments: setting range is 0 ~ 5, it can stop to do magnetic absorption for
each segment, magnetization function will be closed if set it to 0.
Cycle times: repeat magnetic absorption times.
Mag.speed: Its magnetic absorption speed when magnetic rod moves
under the liquid level. 1 is the slowest while 10 the fastest.
Lip lvl: the standing time when magnetic rods closing to liquid level after
finishing magnetic absorption which is for magnetic beads gathering in
case beads falling off due to liquid surface tension.
Anti splash: the standing time when magnetic rods pulling away from
liquid level after finishing magnetic absorption, in case cross
contamination which caused by liquid splashing due to some special
sample tissues falling off.
1 5 Segment time: independent magnetic absorption time of each
segment, the maximum time can reach to 999 seconds.
Estimated: The estimated magnetic absorption time of the software.It can
only be displayed on the next entry after exiting the interface.
3.3.3. Option
In program new or edit interface, click the Option to enter the
option interface. The parameters in the option are applied to the whole
program as shown in the figure below.