8
Plumbing
Flow Rate:
½ gallons per minute (1.9 L/minute)
Minimum Inlet Water Pressure:
25 psi (1.76 kgf/cm²)
Maximum Inlet Water Pressure:
50 psi (3.52 kgf/cm²)
Drainage:
1 gal/minute (3.8 L/minute) (recommended @ 30 psi)
Inlet Tubing:
¼" OD (6.35 mm)
Drainage Tubing:
1¼" ID (32 mm)
Water Inlet
Install the supplied inlet water filter between the water source and the MultiPrep™ base.
The inlet fitting on the rear panel of the MultiPrep™ base (Figure 1) only requires the tube to be
inserted until it stops. No tools are required for this installation.
To remove the tubing, press the outer collar of the fitting in and pull the tubing out.
Note:
Turn the water off at the source before removing the water tubing.
Drain
With the hose clamp on the drainage hose, position the hose onto the coupling at the rear service
panel (Figure 1). Tighten the hose clamp so the hose does not slip from the coupling. Use the 90
degree elbow fitting to direct the drain hose if the tubing does not bend sharply enough for proper
installation.
Activate the coolant solenoid by pressing the coolant button (water flow may need to be regulated
or adjusted using the valve).
Note
: Do not use the valve at the water spout to completely shut off the water flow. It is rated
for 30 psi and is only intended to restrict the water flow during polishing.
Summary of Contents for MultiPrep System 15-2000-GI
Page 35: ...35 MultiPrep System Wiring Diagrams ...
Page 36: ...36 ...
Page 37: ...37 ...
Page 38: ...38 ...
Page 39: ...39 CE Certificates ...
Page 40: ...40 ...
Page 41: ...41 ...
Page 42: ...42 This page has been left blank intentionally ...
Page 43: ...43 This page has been left blank intentionally ...