![allen A-71101-EU User Manual Download Page 56](http://html.mh-extra.com/html/allen/a-71101-eu/a-71101-eu_user-manual_2898490056.webp)
Specifications
56
Allen® Advance Table User Guide (D-720722 Ver D)
Classification
Standard
IEC 60601-1 Classification
According to Degree
Protection against Electrical
Shock
Class I equipment
IEC 60601-1 Classification
According to Degree
Protection against Electrical
Shock
Type B applied part
IEC 60601-2-46 Classification
According to Degree of
Protection against Dust and
Fluid Intrusion
IPX 4
Mode of Operation
2 minutes ON/18 minutes OFF
Applied parts (in accordance with IEC 60601-1)
Rail top
Flat top
Positioning supports (such as pads,
C-Flex® Head Positioning System,
and disposables)
Flat Top pad
Electromagnetic Emissions Guidance
WARNING:
If the Allen® Advance Table is to be used near high-frequency surgical
equipment, cardiac defibrillators, and cardiac defibrillator-monitors,
refer to the product manufacturer’s instructions to make sure the
equipment will operate as intended.
This device meets all requirements for electromagnetic compatibility per
IEC 60601-1-2. Other products that are used near the Allen® Advance
Table should comply with this standard also. Otherwise, electromagnetic
interference between the products could cause the products to operate
incorrectly. If problems do occur, contact the product manufacturer(s).
WARNING:
The A-71100/A-71101 should not be used adjacent to other electrical
equipment. If adjacent use is necessary, observe the A-71100/A-71101
and the other electrical equipment to make sure they operate as
intended.
CAUTION:
This device meets all requirements for electromagnetic compatibility
per IEC 60601-1-2. It is unlikely that the user will encounter problems
with this device because of inadequate electromagnetic immunity.