Recommended separation distances between
portable and mobile RF communications equipment and KardiaMobile® (Model AC-
019)
KardiaMobile® (Model AC-019) is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of KardiaMobile® (Model AC-019)
can help prevent electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and KardiaMobile® (Model AC-019) as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80
MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1—At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2—These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
P
V
d
]
5
.
3
[
1
P
E
d
]
5
.
3
[
1
P
E
d
]
7
[