Instructions for Use for KardiaMobile® (Model AC-019)
Page 16 of 23
CAUTION:
Changes or modifications not expressly approved by AliveCor could void your
authority to use this equipment.
To view FCC information on the Kardia app:
1. On the home screen, tap
to access Kardia app Settings.
2. Tap “About Kardia” to the FCC ID and other applicable regulatory information.
Ingress Protection Marking
KardiaMobile® (Model AC-019) is IP22 rated. KardiaMobile® (Model AC-019) is protected
against insertion of fingers and is not affected by vertically dripping water. KardiaMobile®
(Model AC-019) has been tested with relevant requirement standard IEC 60601-1-11:2015.
Applied Parts
The 3 electrodes (Left Hand Electrode, Right Hand Electrode, and Left Leg Electrode) are Type
CF Applied Parts.
Operational temperature of the device is +10°C to +45°C. If ambient temperature exceeds
+41°C, Applied Parts can 41°C.
Troubleshooting
If you experience difficulties in operating your KardiaMobile® (Model AC-019) System, refer to
the troubleshooting guide below or contact technical support at [email protected].
1. Problem: My KardiaMobile® (Model AC-019) is not working.
Solution
○ Option 1: Make sure you have Bluetooth turned on your smartphone (or tablet).
○ Option 2: Change the battery. AliveCor recommends that you bring your
KardiaMobile® (Model AC-019) hardware to a battery replacement shop for battery
replacement.
2. Problem: I have a lot of artifact, noise, or interference in my recording.
Solution
○ Option 1: Ensure that the “Enhanced Filter” is on.
○ Option 2: Ensure that your arms, hands, and left leg remain still during recordings to
reduce muscle noise.
○ Option 3: Clean the electrodes on the KardiaMobile® (Model AC-019) by spraying
with an alcohol-based sanitizer and wiping with a soft cloth.
○ Option 4: If your hands or left leg are very dry, use a water-based lotion before
recording.