6.2
Status LED Lights
The Centurion status can also be assessed by viewing the status LEDs on the main circuit
board while the unit is on. Only STAT4 and STAT5 LEDs display status. STAT4 displays the Log
Export State and STAT5 displays the Centurion State.
The LEDs various displays indicate the following:
STAT4:
o
LED Off - Log save not in progress
o
LED On - Log save triggered or in progress
STAT5:
o
LED fast blinking – Reset
o
LED two short pulses – Starting up
o
LED continuous one short pulse – Running
o
LED steady blinking – Shutting down
o
LED on – Shut down (power still connected and on)
o
LED off – Unit Off
If any other blinking patterns are noticed, contact your authorised Alcolizer service
technician on 1300 789 908.
6.3
Keyboard
A standard 101 style keyboard (with a standard USB connection) or a wireless keyboard can
be connected to the Centurion to access certain menus during setup and fault finding. Users
should not access the machine and attach a keyboard unless specifically following
instructions in this manual or verbal instructions when in contact with your Alcolizer
Technology technical representative.
Access the USB ports by unlocking the Centurion and swinging it gently open. There are four
ports located on the main circuit panel.
WARNING
When the instrument is open or not installed on the back plate, electrical components are
exposed. The following steps require the user to open the Centurion with power applied.
Do not come in contact with any exposed components.
CAUTION
Use care when inserting the connector to the port to ensure neither are damaged.