2.2.5. Adjusting spring tension to the intensity
required by the patient
1.
Through the window in the PCL casing the set spring force is indicated on a
scale from 0 to 15. The ranges above 15 and below 0 are marked in red.
To prevent damage to the PCL hinge, it must not be rotated into the
red zone of the window on the PCL hinge.
2.
Insert the tool into the lateral hole in the hinge as far as it will go. Turning
clockwise or in the + direction increases spring tension whilst turning
anticlockwise or in the - direction reduces spring tension.
1
2
2.2.6. Altering spring tension
Spring tension can be adjusted according to the progress of treatment.
Insert the tool into the lateral hole in the hinge as far as it will go. Turning
clockwise or in the + direction increases spring tension whilst turning
anticlockwise or in the - direction reduces spring tension.
The intensity of spring tension must be identical on both hinges.
Adjustment of spring force may only be performed in agreement with
the attending physician.
20
Jack PCL Brace
User Instructions