5.3.4 Fault indication in the event of excessive temperatures
LED display
Number of
sounds
Sound
interval
(seconds)
Failure description
Remedy
●○○●○
1
10
High temperature at circuit board
Switch the system off and wait 60 minutes.
Check the ambient temperature.
If the fault occurs repeatedly after switch-
ing on again:
the system needs to be repaired, contact
your specialist dealer.
●○○●○
2
10
High temperature of the battery
5.3.5 General faults
LED display
Number of
sounds
Sound
interval
(seconds)
Failure description
Remedy
●○○○●
2
---
Internal fault, the system is defective. Contact your specialist dealer.
If any fault indications other than those listed are indicated, there is a serious system malfunction. In this case
the twion wheel cannot be used and must be sent to Ulrich Alber GmbH via your specialist dealer for testing.
6. Automatic shut-down
If a wheel that has been switched off does not move for longer than 15 minutes, it is automatically turned off to save energy.
7. Replacing the fuse
As with all electrical and electronic devices, your twion is equipped with a fuse system. In the event of a defect, the fuse reacts and
thereby protects the electronics from sustained and expensive damage.
The fuse of the twion wheel is located within the electronics housed in the wheel hub. Any fuse replacement required, combined with a full
diagnosis of the entire electronic system, may only be performed by Alber.
!