Copyright © 2020
Alazar Technologies Inc.
66
ATS9353 User Manual
Dynamic Parameters
Typical values measured on the 400 mV range of CH A of a randomly selected ATS9353.
Input signal was provided by a Rohde & Schwarz SMB100A signal generator, followed by a
9-pole, 10 MHz band-pass filter (TTE Q36T-10M-1M-50-720BMF). Input frequency was set
at 9.9 MHz and output amplitude was 270 mV rms, which was approximately 95% of the
full scale input.
SNR
60.26 dB
SINAD
59.22 dB
THD
-65.95 dB
SFDR
69.76 dBc
Note that these dynamic parameters may vary from one unit to another, with input
frequency and with the full scale input range selected.
ECLK (External Clock) Input
Signal level
250 mV
p-p
to 2 V
p-p
Input impedance
50 Ω
Input coupling
AC
Maximum frequency
500 MHz for Fast External Clock
Minimum frequency
25 MHz for Fast External Clock
Sampling edge
Rising
Optional 10 MHz Reference PLL Input
Signal level
200 mV
p-p
to 2 V
p-p
Input impedance
50 Ω
Input coupling
AC
Input frequency
10 MHz ± 0.1 MHz
Maximum frequency
10.1 MHz
Minimum frequency
9.9 MHz
Sampling clock freq.
500 MHz