* The battery of AKTIIA INIT I1 is built-in rechargeable lithium-ion battery, please do not disassemble it by the unauthorized maintenance personel.
* Under normal usage, it can charge power about 300 times, if the battery cannot charge the power normally or the blood pressure monitor cannot be used normally,
please connect with the authorized maintenance personel.If measured three times per day, and the battery is fully charged, it can be used for about 23 days.
* Store and use the blood pressure monitor in a cool, dry and ventilated environment. Avoid to approach to fire and heat sources or the battery may explode.
* Only AKTIIA's authorized AC Adaptor can be used ( Model: BLJ06L050100U-V) (Not included) to
charge
the
power.
You cannot use the blood pressure monitor
during the process of charging.
* During the process of charging, the light turns green, it means the charging is finished, please pull the plug.
* When charging, shall not touch charging connector and the patient simultaneously.
* Do not attempt to replace your blood pressure monitor’s battery. It is built-in and not changeable.
* Only charge the battery in accordance with the user instructions supplied with the blood pressure monitor.
* Avoid charging your blood pressure monitor in extremely high or low temperatures.
* Do not use your blood pressure monitor while you are charging it.
* Do not attempt to disassemble the blood pressure monitor or force open the built-in battery.
* Do not clean the blood pressure monitor when it is being charged. Always unplug the charger first before cleaning the blood pressure monitor.
* Do not dispose of your blood pressure monitor in a fire. The battery could explode causing injury or death.
* Batteries (battery pack or batteries installed) shall not be exposed to excessive heat such as sunshine, fire or the like.
Charge at least once every three months. If the battery loses all its charge, it may not be rechargeable.
PROBLEM
SYMPTOM
CHECK THIS
REMEDY
No power
Low
Measurement error
batteries
The light is not on.
The battery is empty
Recharge the battery.
The battery is low.
The button is in the “OFF” side.
Switch the button on the “ON” side.
Red light flashing.
Recharge the battery.
APP displays "Recording failed".
12
Cuff not tight or inflated properly, talk or walk while
measuring and the measurement is out of range.
Adjust the cuff, hold still, and
measure again.