Setting procedure: turn on power, rotate counterclockwise the CC 

adjustment (2) and (17) to minimum, then make the positive and 

negative output terminal in short connection and rotate clockwise 

the CC adjustment (2) and (17) till output current equal to the 

required current-limiting protection point, so the current-limiting 

protection point is well set. 

2.2.2 Series using of the two adjustable outputs 

2.2.2.1 Switch (9) is set to spring out and press in switch (7). At this 

time, turn the master voltage adjustment (14) and the slave out 

voltage tracks strictly the master output voltage, and the output 

voltage can be up to double of independent's maximum voltage 

(voltage between terminal (18) and (23)). 

2.2.2.2 Before the series connecting, it must be examined if the 

negative terminal of both master and slave output are connected to 

case grounded terminal, if they are, must be disconnect, otherwise, 

short-circuit will be caused in the slave output when the two outputs 

are connected is series. 

2.2.2.3 When the two outputs are in series, the voltage is controlled 

by master output, but current adjustment of two outputs is still 

independent. Therefore, attention should be paid to the position of 

the CC adjustment (2). For example, knob (2) is at the position of 

counterclockwise to end or current of slave output excesses current-

limiting protection point, at this time, the voltage of slave output will 

not track the voltage of master. So knob (2) should be rotated 

clockwise to maximum then the two outputs are in series. 

2.2.2.4 By series connection, if there is power output, proper leads 

corresponding to output power should be used to short connect the 

negative terminal of master output with positive terminal of slave 

output reliably. Since it is shorted by a switch inside the unit, current 

will pass on the shorted switch when there is power output. This will 

affect the reliability of the unit. 

2.2.3 Parallel using of the two adjustable outputs 

2.2.3.1 Press in switch (9) as well as switch (7), at this time, the two 

output are in parallel, adjust voltage adjustment (14) of master 

output, the voltage of two ways keep same, and slave output CC 

indicator (1) lights on. 

Summary of Contents for ATH-2235

Page 1: ...User s Manual DC POWER SUPPLIES ATH 2235 ATH 2243 ATH 3231 ATH 3232 ATH 4012 Thank you for choosing AKTAKOM Power Supply www tmatlantic com...

Page 2: ...uipment to direct sunlight or extreme temperatures Do not subject the equipment to extreme humidity or dampness Replace a defective fuse only with a fuse of the original rating Never short circuit fus...

Page 3: ...lies have three ways output Of which two ways are adjustable and one way is fixed The two adjustable outputs can also be selected for constant voltage or constant current designed in high stability an...

Page 4: ...gulation Two adjustable output CV 1 x 10 4 3mV CC 2 x 10 3 3mA Fixed output 10mV 1 4 Load regulation Two adjustable outputs CV 1 x 10 4 2 mV I 3 A CV 1 x 10 4 5 mV I 3 A CC 2 x 10 3 3 mA I 3 A CC 2 x...

Page 5: ...ng slave output current by analog meter or LED LCD 6 Volt display indicating slave output voltage by analog meter or LED LCD 7 Control switch for selecting the two adjustable outputs independent serie...

Page 6: ...put 2 2 1 1 Set 7 and 9 switch to spring out position 2 2 1 2 When the adjustable output is used as CV output first should rotate clockwise the CC adjustment 2 and 17 to maximum then turn on power swi...

Page 7: ...onnected is series 2 2 2 3 When the two outputs are in series the voltage is controlled by master output but current adjustment of two outputs is still independent Therefore attention should be paid t...

Page 8: ...ter is 2 5 class To get more accurate measuring value you should calibrate by external circuit with precision measuring instrument 3 CAUTION 3 1 This unit has excellent protection function 5V output h...

Reviews: