ABM-4083
90
:AUTO <b>
Command Syntax:
:TRIGger:DELay:AUTO <b>
Command parameter:
<b> = 1 or ON
Enable atuo delay
0 or OFF
Disable atuo delay
Query:
:AUTO?
Query delay state
Description:
This command is used for enable or disable auto delay state at trigger model.
:COUNt <n>
Command Syntax:
:TRIGger:COUNt <n>
Command parameter:
<n> = 1 to 9999
Set trigger count
INFinite
Set trigger count of infinite
MINimum
Set trigger count of 1
MAXimum
Set trigger count of 9999
Query:
COUNt?
Query trigger count
Description:
Specify the trigger count before the digital multimeter go back to idle. Parameter INFinite indicates the
digital multimeter can receive trigger continuously. When the instrument is at local control, the trigger
count is ignored. The reference value is saved in the volatile memory.
:SAMPle
:COUNt <NRf>
Command Syntax:
:SAMPle:COUNt <NRf>
Command parameter:
<NRf> =
1 to 30000
Set sample count
MINimum
Set sample count of 1
MAXimum
Set sample count of 30000
Query:
:SAMPle?
Description: