ABM-4083
80
[:SENSe[1]]:PERiod:DIGits <n> Specify resolution for PER
Command parameter:
<n> = 4 3
1
/
2
digits
5 4
1
/
2
digits
6 5
1
/
2
digits
7 6
1
/
2
digits
DEFault 5
1
/
2
digits
MINimum 3
1
/
2
digits
MAXimum 6
1
/
2
digits
Query:
DIGits? Query selected resolution
Description:
These Commands are used to select display resolution for the specified measurement function.
Even though the parameters for this command are expressed as integers(4 to 7),you can specify res-
olution using real numbers.For example,to select 3
1
/
2
digit resolution let <n>= 3.5,for 4
1
/
2
digit resolution let
<n>= 4.5,and so on.
:AVERage Commands
This command is used to configure and control the filter.
:STATe <b>
Command syntax:
[:SENSe[1]]:CURRent:AC:AVERage:STATe <b>
Control filter for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:STATe <b>
Control filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:STATe <b>
Control filter for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:STATe <b>
Control filter for DCV
[:SENSe[1]]:RESistance:AVERage:STATe <b>
Control filter for
Ω2
[:SENSe[1]]:FRESistance:AVERage:STATe <b>
Control filter for
Ω4
Command parameter:
<b> = 1 or ON
Enable the digital filter
0 or OFF
Disable the digital filter
Query:
:STATe?
Query state of digital filter
Description:
These commands are used to enable or disable the digital filter for the specified function. When ena-
bled, readings will be filtered according to how the filter is configured.
:TCONtrol <name>
Command syntax:
:CURRent:AC:AVERage:TCONtrol <name>
Select filter type for ACI