ABM-4083
89
Description:
When continuous initiation is selected (ON), the instrument is taken out of the idle state. At the con-
clusion of all programmed operations, the instrument returns to the top of the trigger model.
NOTE: With continuous initiation enabled (ON), you cannot use the :READ? command or set
sample count greater than one (see:SAMPle:COUNt).
:ABORt Command
:ABORt
Description:
When this action command is sent, Model DMM aborts operation and returns to the top of the Trigger
Model.
:TRIGger
TRIGger subsystem commands are used for the instrument settings of trigger mode, trigger delay and
trigger a measurement.
:SOURce <name>
Command Syntax:
TRIGger:SOURce <name>
Command parameter:
<name> = IMMediate
instrument default setting is internal trigger)
BUS Triggered via RS232 interface of GPIB interface)
MANual(EXTernal) (Press
Trig
in the front panel for Trigger)
Query:
:SOURce?
Query control source)
Description:
These are used to select the event control source.
:DELay <n>
Command Syntax:
:TRIGger:DELay <n>
Command parameter:
<n> = 0 to 60000
Specify delay in
MINimum
Specify delay of 0S
MAXimum
Specify delay of 60S
Query:
:DELay?
Query the programmed delay
Description:
This command is used for delay operation of the trigger model. The delay time can also be set by using
the AUTO parameter and this value is stored in the volatile memory.