7.4 Replace the air filter
Air filter is at the back of the device, remove it by
tunning it anti-
clockwise and install
a new one instead.
7.5 Door adjustment
On normal circumstance the chamber door lock don't need to adjust.Once steam
leaking occurs (the seal fails), you may use the spanner to adjust door seal.
7.51
Open the door f
irstly.
7.5.2
Insert the spanner into the gap beneath the plastic cover; use the spanner to
lock on the adjusting nut (Fig 1). Turn the nut counter clockwise as the figure below
(Fig 2). This will tighten the sealing plate.
7.5.3
Turn the nut until the sea
ling plate is tight. If the door knob is too tight, you
may also turn the nut clockwise to loosen it.
Caution: