Instruction manual
- 13 -
TEMP button to adjust, press START button to confirm it.
Please set the parameter as following:
Altitude Parameter Altitude Parameter
1.0>h>0.5 km, 0.5; 1.5>h>1.0 km, 1.0;
2.0>h>1.5 km, 1.5; 2.5>h>2.0 km, 2.0;
3.0>h>2.5 km, 2.5; 3.5>h>3.0 km, 3.0;
4.0>h>3.5 km, 3.5; 4.5>h>4.0 km, 4.0;
Note:
The standard atmospheric pressure is about 100kPa. And the pressure decrease 5kPa for
each 0.5 kilometers of altitude increased.
If the parameter is set above 2, you need to reevaluate the sterilization result.
And you may correct the effect by prolong the holding time.
6.3.3
Print and USB setting
‘ 1’
means it is effective ,
‘0’means it is noneffective. The left number refers to
USB, and the right one refers to print function.
6.3.4 Identification code
You shall not change the identification
code of your device, it i
s used for
maintenance trace.