Set default:
This option allows you to save all of the monitor’s variable settings; these settings will then remain active even when
the monitor is rebooted.
The following settings can be stored:
Printer ON/OFF, Alarm ON/OFF, Measurement settings A/B/C/D/E/off, SMS ON/OFF, NAME, SMS delay, APN,
Internet ON/OFF, PIN code modem, SMS tel. no., P1 Relay, SND delay, Language, Sensors, High level E, Low level E.
Please note!
This feature is only available if the ‘Limited menu’ feature has been turned off.
1.
Press the Menu key 9
to open the menu.
2.
Press key 12 3 times to select the SYSTEM menu.
3.
Press the Enter key 13 to open the SYSTEM menu.
4.
Press key 11 until the ‘Set default’ option is shown.
5.
Press the Enter key 13 : the current settings will be saved as the default setting and the monitor will return
to its home screen.
Language:
This feature allows you to set the monitor’s language. The monitor is able to utilise the following languages:
Dutch/German/French/English.
1.
Press the Menu key 9
to open the menu.
2.
Press key 12 3 times to select the SYSTEM menu.
3.
Press the Enter key 13 to open the SYSTEM menu.
4.
Press key 11 until the ‘Language’ option is shown.
5.
Press the Enter key 13 .
6.
Press keys 10 and/or Nr.11 to select the desired language.
7.
Press the Enter key 13 to confirm your choice.
8.
Press the Menu key 9
2 times to return to the home screen.
Settings:
Option of configuring the monitor.
Please note!
This feature is only available if the ‘Limited menu’ feature has been turned off.
This feature is only intended for Airtocheck service staff.