SPECIFICATION

DESCRIPTION

Sampling Method 

Passive radon diffusion chamber 

Alpha spectrometry

Detection Method

4 silicon photodiodes in 4 

distinct radon chambers

Detector Power Supply

3 AA alkaline batteries (LR06)  

~ 15-month battery life

Dimensions

5.5 x 5.5 x 1.2 inches 140 x 140 

x 30 mm

Weight

10.6 ounces (including 

batteries)  

300 grams (including batteries)

Operation Environment

39 to 104 °F 

4 to 40 °C 

5% RH to 85% RH noncondensing 

50 kPa to 110 kPa

Measurement Range 

0 - 2700 pCi/L 

0 - 100 000 Bq/m

3

Measurement 

Uncertainty 

After 24 hours  

After 7 days

σ

 ~ 7 % ± 0.15 pCi/L  

σ

 ~ 7 % ± 5 Bq/m

3

 

σ

~ 5 % ± 0.06 pCi/L  

σ

 ~ 5 % ± 2 Bq/m

3

Diffusion Time Constant 

25 minutes

Temperature Sensor 

Range  

Resolution  

Accuracy

39 to 104 °F  

4 to 40 °C  

0.36 °F 

0.2 °C 

± 1 °F (typical), ± 2 °F (max) 

± 0.5 °C (typical), ± 1 °C (max)

Humidity Sensor 

Range  

Resolution  

Accuracy

5% RH to 85% RH (non-

condensing) 0.5% RH 

± 4.5% (in range 20 - 85% RH)

SPECIFICATION

DESCRIPTION

Barometric Pressure 

Sensor 

Range  

Resolution  

Accuracy

50.0 kPa to 115.0 kPa 

0.06 kPa 

± 1 kPa

Memory

The internal memory stores 5 

years of hourly datapoints

Memory Capacity

~ 1 900 days of measurements 

~ 5 datasets of one-year length  

~ 177 datasets of 1-week 

length  

~ 325 datasets of 2-day length

Memory Type

 Non-volatile flash memory

SPECIFICATIONS

USAGE

USAGE & SAFETY

The radon sensors in Corentium Pro are made for 

inside use only. Avoid storing or using the monitor 

in environments with high humidity, extreme heat, or 

freezing conditions. Condensation can form on the 

monitor if it is moved from a warm location to a cold 

location or vice versa. If condensation forms on the 

monitor, remove the batteries and leave it in a dry 

environment for 2 hours. The monitor should not be 

used in environments with more than 85% relative 

humidity. The unit must not be exposed to direct sunlight 

for extended periods. Avoid subjecting the monitor to 

shocks, impacts, pressure, vibrations, dust, or moisture.

Only use alkaline AA (LR06) batteries. The batteries must 

not be exposed to fire or other extreme heat. The battery 

terminals must not be touched, and they must be kept 

free from dust, sand, liquids and other foreign objects.

Use a Kensington cable Lock (often used for laptops) 

Summary of Contents for Corentium Pro

Page 1: ...f fresh air A Reset button B Kensington Security Slot C Mini USB connection There is no ON OFF button 1 Communication status light 2 Status lights LEDs 3 Battery compartment 4 Tripod mount 5 Serial number label 6 Calibration label A B C 1 3 4 5 6 2 ...

Page 2: ... kPa 1 kPa Memory The internal memory stores 5 years of hourly datapoints Memory Capacity 1 900 days of measurements 5 datasets of one year length 177 datasets of 1 week length 325 datasets of 2 day length Memory Type Non volatile flash memory SPECIFICATIONS USAGE USAGE SAFETY The radon sensors in Corentium Pro are made for inside use only Avoid storing or using the monitor in environments with hi...

Page 3: ...ection CREATING AN AIRTHINGS ACCOUNT In order to use the mobile app and online Pro Dashboard a user must create an Airthings account by visiting the website at pro airthings com and clicking on the Sign Up with Airthings button Follow the remaining prompts to create an account USING THE ONLINE PRO DASHBOARD Radon tests are uploaded from the mobile app to the Dashboard and can be accessed from any ...

Page 4: ...ready for use click on Publish You may also use the duplicate feature to create additional templates Generating a Report in the Dashboard Once a radon test dataset has been uploaded to the Dashboard it can be accessed by clicking on the datasets icon in the left menu A list of all datasets in the Dashboard is presented and a specific dataset can be located using the search field To generate a repo...

Page 5: ...adon Test To start a radon test connect to a monitor as described in the Connecting to Corentium Pro section If the monitor is available to use for a radon test a green button that says Start Test will be seen at the bottom of the screen tap on this button to begin the scheduling process You will be prompted to enter some mandatory information while the rest are optional Airthings recommends a min...

Page 6: ...ersion date and time calibration expiration date or status of a monitor connect to it as described in the section above Once connected to the monitor tap on the i button near the serial number at the top of the screen Other Operations To log out of the app from the home screen tap on the gear icon at the top right corner of the screen then tap on Log Out You can also get help with the Corentium Pr...

Page 7: ...rsuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residual installation This equipment generates uses and can radiate frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in ...