pNeuton Ventilator
8 - 3
Indication Meaning
Corrective
Action
Patient pressure
too high
Tidal Volume set too
high
Decrease Tidal
Volume or Peak
Pressure setting
Patient response
ET-Tube may be
occluded or patient
may be biting tube
Expiratory
Valve
malfunctioning
Replace patient
circuit
Internal malfunction
Send ventilator for
service
Can’t get the
PEEP / CPAP
desired
Expiratory Valve
malfunctioning
Replace patient
circuit
Using a circuit not
recommended by
Airon
Replace patient
circuit
Internal malfunction
Send ventilator for
service
Excessive
“chattering” of CPAP
system
Occurs when using
some test lungs but
will not when
connected to a
patient. If problem
persists, send
ventilator for service
Ventilator using
too much gas
PEEP / CPAP
system turned “on”
Turn off PEEP /
CPAP system
Leak at source gas
Check hoses and
tank regulator for
leaks
Internal leaks
Send ventilator for
service
Oxygen
concentration too
low
Source gas not
100% oxygen
Insure source gas is
100% oxygen
High
patient
spontaneous
ventilation
Decrease
spontaneous
ventilation
Internal malfunction
Send ventilator for
service
Summary of Contents for pNeuton A
Page 1: ...pNeuton Transport Ventilator Model A Operators Manual 97011 CD A 005 Rev J Released 1 Feb 2019...
Page 6: ...pNeuton Ventilator 1 2...
Page 14: ...pNeuton Ventilator 2 8...
Page 22: ...pNeuton Ventilator 3 8...
Page 32: ...pNeuton Ventilator 5 4...
Page 36: ...pNeuton Ventilator 6 4...
Page 58: ...pNeuton Ventilator 10 4...