pNeuton Ventilator
8 - 1
Section 8:
Troubleshooting
This troubleshooting guide lists common problems that
may be encountered and possible solutions. If none of the
corrective actions seem to work, contact Airon Corporation
or your distributor.
Indication Meaning
Corrective
Action
Ventilator does
not operate – no
patient ventilation
Missing or
insufficient driving
gas supply
Check gas source,
55 psi (380 kPa) at
40 L/min is required
Patient
circuit
disconnection
Reconnect patient
circuit
Internal malfunction
Send ventilator for
service
Ventilator seems
to “want” to
operate, but no
breaths are
generated
Peak Pressure
control set too low
Increase Peak
Pressure control
Respiratory Rate set
too low
Increase Respiratory
Rate
Expiratory
valve
drive line
disconnected
Insure tubing is
properly connected
Expiratory Valve is
malfunctioning
Replace patient
circuit
Insufficient
driving
gas supply
Check gas source,
55 psi (380 kPa) at
40 L/min is required
Internal malfunction
Send ventilator for
service
Ventilator
appears to be
stuck in
inspiration
CPAP may be
turned on high
Check CPAP control
Internal malfunction
Send ventilator for
service
Ventilator stops
and starts
Insufficient driving
gas supply
Check gas source,
55 psi (380 kPa) at
40 L/min is required
Summary of Contents for pNeuton A
Page 1: ...pNeuton Transport Ventilator Model A Operators Manual 97011 CD A 005 Rev J Released 1 Feb 2019...
Page 6: ...pNeuton Ventilator 1 2...
Page 14: ...pNeuton Ventilator 2 8...
Page 22: ...pNeuton Ventilator 3 8...
Page 32: ...pNeuton Ventilator 5 4...
Page 36: ...pNeuton Ventilator 6 4...
Page 58: ...pNeuton Ventilator 10 4...