![Airon pNeuton A Operator'S Manual Download Page 50](http://html1.mh-extra.com/html/airon/pneuton-a/pneuton-a_operators-manual_2878642050.webp)
pNeuton Ventilator
8 - 2
Indication Meaning
Corrective
Action
Lower minute
volume than
desired
Insufficient driving
gas supply
Check gas source,
55 psi (380 kPa) at
40 L/min is required
Leak in the Patient
Circuit or Expiratory
Valve
Replace patient
circuit
Obstruction of gas
output
Check or replace
patient circuit
Use in hyperbaric
condition
Ventilator should not
be used in
hyperbaric conditions
Tidal
volume
control
out of calibration
Send ventilator for
service
Internal malfunction
Send ventilator for
service
Higher minute
volume then
desired
Use at higher
altitude then
calibration
Use external
spirometer to verify
tidal volume
Tidal
volume
control
out of calibration
Send ventilator for
service
Internal malfunction
Send ventilator for
service
Tidal volume
inaccurate
Leak in the patient
ET-Tube, mask,
breathing circuit or
expiratory valve
Check patient
interface. Replace
patient circuit if at
fault
Ventilator
is
operating at an
altitude different
then calibration
Tidal volume should
be measured by an
external spirometer
Tidal
volume
control
out of cal
Send ventilator for
service
Rate control
inaccurate
Tidal volume set
below 500 ml or
above 900 ml
This is normal. Rate
will be faster when
tidal volume is set
lower than 500 ml.
Rate will be slower
when tidal volume is
set higher than 900
ml
Rate control out of
cal
Send ventilator for
service
Summary of Contents for pNeuton A
Page 1: ...pNeuton Transport Ventilator Model A Operators Manual 97011 CD A 005 Rev J Released 1 Feb 2019...
Page 6: ...pNeuton Ventilator 1 2...
Page 14: ...pNeuton Ventilator 2 8...
Page 22: ...pNeuton Ventilator 3 8...
Page 32: ...pNeuton Ventilator 5 4...
Page 36: ...pNeuton Ventilator 6 4...
Page 58: ...pNeuton Ventilator 10 4...