User manual
–
HEPA
Bulle
–
ED02_09-August-2021
14 /31
2.2
START-UP
Connect the
HEPA
Bulle
unit to an electrical socket ~ 230 V
– 50/60 Hz using the 2P+T 10 A power
supply cable provided. The screen turns on.
IMPORTANT:
THE ELECTRICAL SUPPLY PLUG MUST COMPLY WITH REGULATORY
TEXTS AND BE PERIODICALLY CHECKED. THE POWER SUPPLY CABLE
MUST BE ACCESSIBLE AT ALL TIMES SO THE PRODUCT CAN BE
UNPLUGGED EASILY, AS NEEDED.
The
HEPA
Bulle
unit loads the operating programme. The initialisation screen appears
(decontamination is not active).
2.2.1 FIRST BOOT
The first time the
HEPA
Bulle
unit boots up, a succession of specific screens appears (called first boot
configuration), where the settings described hereafter must be adjusted. On subsequent start-ups, the
main screen will open directly (see §2.1.1).
NOTE:
Confirm the current screen and move to the next screen by pressing the
icon. You
can return to the previous screen by pressing the
icon.
NOTE:
You can reactivate the display of the first boot procedure in the machine settings (see
§3.1).
2.2.1.1
Interface language selection