Air Techniques PERI-PRO III User Manual Download Page 8

8

MAINTENANCE

Servicing your Peri-Pro III on a daily, two week and quarterly basis is critical to maintaining quality 

processor performance.  We recommend the maintenance procedure outlined below, using Formula 

2000 Plus cleaner, and Peri-Pro Developer and Fixer, which are specially formulated for Peri-Pro pro-

cessors. This maintenance schedule may have to be modified if more frequent cleaning is required.

DAILY - CHANGE WATER

TOP OFF CHEMISTRY TANKS WITH WATER TO FILL LINE

Important:

  Make sure not to mix the developer and fixer chemistry.  

Contaminating the solution will degrade image quality.

  Dispose of wash water in accordance with local codes.

EVERY TWO WEEKS (or every 300 - 350 films) 

SCRUB TRANSPORT; CHANGE PERI-PRO CHEMISTRY

Caution:

  Cleaning and changing chemistry requires the 

removal  of  the  cover  and  film  transport  from 

the Peri-Pro III.  Always turn POWER off and 

unplug the line cord before removing the cover 

or transport.

Scrub Transport

1. Turn 

POWER OFF

 and unplug the line cord. Remove the film receptacle, lift off the cover.

2  Lift the film transport and let the solutions drip back into the tanks before placing the transport 

in a sink. (Use of optional Service Tray PN 90121 will contain any dripping chemistry.)

3.  Rinse the transport thoroughly with warm water to soften residue and dried chemistry. Be sure 

to rinse above the chemistry level, keeping the shutter and the film inlet on the top of the 

transport  dry. (Use a sink or the transport wash tub that comes with the Transport Cleaning 

Kit PN 43975.)

1. Turn 

POWER OFF

 and unplug the power cord from the outlet.

2. Remove the film delivery receptacle (see figure 1) at the film dryer exit on the right side of the 

cover.  Remove the main cover by carefully lifting away from the unit.

Important:

  Do not splash chemistry or contamination may result.

3.  As shown by Figure 4, carefully lift the Film Transport and hold over the tanks for a few seconds 

so liquids drain back into tanks and place the transport on the service tray P/N 90121.

4.  Remove the water tank (white), discard the water and rinse out the tank.  If the wash tank is 

particularly soiled or slimy, it must be thoroughly cleaned and rinsed.

5.  Refill the clean water tank with 70° 80°F water; reinsert into base.
6.  Check that the solution in the developer (black) and fixer (red) tanks are up to the Fill Lines as indi-

cated on the inside of the refill access door. If not, top off the developer and fixer tanks with water.

7.  Slowly and carefully lower the Film Transport straight into the tanks. The Film Transport must 

sit squarely on the cutouts found on the top of the vertical plates to the left and right side of 
the tanks. 

8.  Align the Main cover and carefully install the Main Cover onto the unit.
9.  Hang the film delivery receptacle (see figure 1) on the film dryer exit on the right side of the cover.     
10. Plug the power cord into the electrical outlet and turn the 

POWER ON

.

Summary of Contents for PERI-PRO III

Page 1: ...PERI PRO III INTRA ORAL FILM PROCESSOR USER S MANUAL...

Page 2: ...i Pro III uses a heating pad system to maintain uniform chemistry bath temperature ensuring the delivery of consistent film quality Additional features include q No plumbing connection required q Slim...

Page 3: ...ING ELEMENT ACCESS PANEL WASH TANK WHITE FIXER TANK RED FILM INLET GRILL FILM INLET SHUTTER COVER ASSEMBLY MAINTENANCE LABEL OPERATOR CONTROLS FILM DELIVERY RECEPTACLE AIR OUTLET GRILL AIR INLET GRILL...

Page 4: ...he base on a flat level stable counter Important The surface must be stable so that chemistry does not splash or spill Locate the thermistor arm above and to the left of the developer black tank Caref...

Page 5: ...POWER SWITCH Amber When the POWER switch is turned ON it will illuminate indicating that there is power to the proces sor and the heating system is on Turn the POWER switch OFF at the end of the day...

Page 6: ...cle when stripping covers off film The trash receptacle is especially useful when using a Daylight Loader Processing 2 Adult Films 2 films can be fed directly into the processor Duplicating films expo...

Page 7: ...at each film on the carrier is in its own groove c Remove the film inlet grill Remember to replace it when processing 2 films Insert the carrier with the film s standing vertically into the film inlet...

Page 8: ...istry level keeping the shutter and the film inlet on the top of the transport dry Use a sink or the transport wash tub that comes with the Transport Cleaning Kit PN 43975 1 Turn POWER OFF and unplug...

Page 9: ...Peri Pro III 1 Slowly and carefully lower the film transport straight into the tanks The film transport must sit squarely on the cutouts found on the top of the vertical plates to the left and right s...

Page 10: ...Maintenance If the wash tank is particularly soiled or slimy it must be thoroughly cleaned and rinsed After cleaning if the tank was removed Figure 13 Figure 14 CAUTION Use in a well ventilated area...

Page 11: ...ted Peri Pro chemistry d Use Peri Pro chemistry to assure film quality and archival quality 3 Film is cloudy or has brownish surface a Fixer is exhausted a Change chemistry after 300 350 film or 2 wee...

Page 12: ...grooves with a soft tooth brush c Timing of the transport arms is out of adjustment c Call your authorized Air Tech niques dealer for service 7 Developer is black has ammonia smell there are black de...

Page 13: ...thorized Air Techniques dealer for service f Defective drive motor f Call your authorized Air Techniques dealer for service g Films reach dryer and staythere broken gear teeth or defective gears or in...

Page 14: ...15 8 Length Width Height Overall 25 9 3 4 8 1 2 With Daylight Loader 25 9 3 4 15 3 4 With Film Duplicator Daylight Loader 25 12 3 4 15 3 4 Add 3 1 2 to length for film receptacle MINIMUM CLEARANCE TO...

Page 15: ...Cleaning Kit Formula 2000 Plus Twin Pack Contains 2 bottles of Formula 2000 Plus 2 color coded sponges 43945 Peri Pro Chemistry Case contains 3 one qt bottles Developer 3 one qt bottles Fixer 90800 Fi...

Page 16: ...nnovator and manufacturer of dental products Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support Whether the need i...

Reviews: