Air Techniques PERI-PRO III User Manual Download Page 10

10

MAINTENANCE

QUARTERLY - CLEAN WITH FORMULA 2000 Plus

Clean the Peri-Pro III Transport with Formula 2000 Plus quarterly or 

whenever a black residue is evident on the transport or in the tanks.

Use the Peri-Pro Transport Cleaning Kit (PN 43975)

Each kit contains:

Cleaning 

tub

Brush and cleaning sponges

Twin pack of Formula 2000 

Plus

CAUTION:

  Use cold water only.

6.  Sprinkle two powder packets of Component 2 evenly across the cleaning tub. Formula 2000 

Plus’s effervescent cleaning action will  begin as soon as Component 2 is added. Some odor 

and vapor will be evident.

7.   Soak the transport for approximately 30 minutes, or until the effervescent action stops.
8.   Remove the transport and, under cold running water, remove any remaining deposits with the 

brush; rinse.

9.   Dispose of the cleaning solution in accordance with local codes. Rinse the cleaning tub and 

place the transport back in the tub. Fill with fresh water and drain. Rinse and drain at least 2 

more times. (Be sure to rinse thoroughly since cleaner residue can contaminate fresh chemistry.)

10. Place the transport on a towel with the gears facing up and air dry thoroughly. Check that the 

shutter area is dry before using. If not, dry with a soft, lint-free towel. See Fig. 14.

11. Change the water. Follow the steps for changing water under Daily Maintenance. If the wash 

tank is particularly soiled or slimy, it must be thoroughly cleaned and rinsed. After cleaning, if 

the tank was removed.

Figure 13

Figure 14

CAUTION: 

Use in a well ventilated area. Wear protective clothing, rubber 

gloves and goggles.  Review the Material Safety Data Sheet 

that comes with every Formula 2000 Plus Twin Pack.

1.  Turn the 

POWER OFF

 and unplug the line cord. Remove the 

film receptacle; lift the cover off.

2.  Remove the transport and rinse under running water.
3.  Place the transport into the cleaning tub with the gears facing 

up    .

4.  Pour two bottles of Formula 2000 Plus Component 1 into the 

cleaning tub. Pour slowly. Do not splash or splatter solution - 

wipe up any spills. (Use both bottles of Formula 2000 Plus in 

the Twin Pack to insure adequate cleaning of the transport.)

5.  Add cold water to the recessed line in the tub. 

12. Clean the Dryer Transport if dry, white chemistry residue is evident on 

the rollers.

 

Remove the screw holding the dryer front cover plate in place.

 

Slide the cover plate up, remove it and lift the dryer transport 

straight up and out.

 

Use warm water and a brush to remove the residue; rinse and 

completely air dry.

 

Reinstall the dryer transport and the cover plate.

Important: 

This plate is necessary for proper film drying.

13. Change the chemistry. Follow the steps for changing chemistry under Every Two Weeks 

Maintenance.

14. Reinstall the transport and reassemble the Peri-Pro III. Follow the steps for reassembly under 

Every Two Weeks Maintenance.

Summary of Contents for PERI-PRO III

Page 1: ...PERI PRO III INTRA ORAL FILM PROCESSOR USER S MANUAL...

Page 2: ...i Pro III uses a heating pad system to maintain uniform chemistry bath temperature ensuring the delivery of consistent film quality Additional features include q No plumbing connection required q Slim...

Page 3: ...ING ELEMENT ACCESS PANEL WASH TANK WHITE FIXER TANK RED FILM INLET GRILL FILM INLET SHUTTER COVER ASSEMBLY MAINTENANCE LABEL OPERATOR CONTROLS FILM DELIVERY RECEPTACLE AIR OUTLET GRILL AIR INLET GRILL...

Page 4: ...he base on a flat level stable counter Important The surface must be stable so that chemistry does not splash or spill Locate the thermistor arm above and to the left of the developer black tank Caref...

Page 5: ...POWER SWITCH Amber When the POWER switch is turned ON it will illuminate indicating that there is power to the proces sor and the heating system is on Turn the POWER switch OFF at the end of the day...

Page 6: ...cle when stripping covers off film The trash receptacle is especially useful when using a Daylight Loader Processing 2 Adult Films 2 films can be fed directly into the processor Duplicating films expo...

Page 7: ...at each film on the carrier is in its own groove c Remove the film inlet grill Remember to replace it when processing 2 films Insert the carrier with the film s standing vertically into the film inlet...

Page 8: ...istry level keeping the shutter and the film inlet on the top of the transport dry Use a sink or the transport wash tub that comes with the Transport Cleaning Kit PN 43975 1 Turn POWER OFF and unplug...

Page 9: ...Peri Pro III 1 Slowly and carefully lower the film transport straight into the tanks The film transport must sit squarely on the cutouts found on the top of the vertical plates to the left and right s...

Page 10: ...Maintenance If the wash tank is particularly soiled or slimy it must be thoroughly cleaned and rinsed After cleaning if the tank was removed Figure 13 Figure 14 CAUTION Use in a well ventilated area...

Page 11: ...ted Peri Pro chemistry d Use Peri Pro chemistry to assure film quality and archival quality 3 Film is cloudy or has brownish surface a Fixer is exhausted a Change chemistry after 300 350 film or 2 wee...

Page 12: ...grooves with a soft tooth brush c Timing of the transport arms is out of adjustment c Call your authorized Air Tech niques dealer for service 7 Developer is black has ammonia smell there are black de...

Page 13: ...thorized Air Techniques dealer for service f Defective drive motor f Call your authorized Air Techniques dealer for service g Films reach dryer and staythere broken gear teeth or defective gears or in...

Page 14: ...15 8 Length Width Height Overall 25 9 3 4 8 1 2 With Daylight Loader 25 9 3 4 15 3 4 With Film Duplicator Daylight Loader 25 12 3 4 15 3 4 Add 3 1 2 to length for film receptacle MINIMUM CLEARANCE TO...

Page 15: ...Cleaning Kit Formula 2000 Plus Twin Pack Contains 2 bottles of Formula 2000 Plus 2 color coded sponges 43945 Peri Pro Chemistry Case contains 3 one qt bottles Developer 3 one qt bottles Fixer 90800 Fi...

Page 16: ...nnovator and manufacturer of dental products Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support Whether the need i...

Reviews: