Air Techniques, Inc
.
Page 6
Product description
3.1 Technical data
Unit electrical data
Nominal voltage
100 to 240 V AC
Max . voltage fluctuation
±10%
Frequency
50/60 Hz
Power rating
500 W
Maximum power
750 VA
Classification
FDA 21 CFR Device Classification
Class IIb
This X-ray system complies with US - FDA:
21 CFR Part 1010 .2
21 CFR Part 1020 .30/31
Degree of protection against ingress of water
Ordinary
Medical products directive (93/42/EEC)
Manufacturer: VATECH Co ., Ltd .
13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, Korea
X-ray tube electrical data
X-ray Tube Type
Canon D-041SB (Stationary Anode Type)
Model
DG-10A05T3
Generator size and weight
5 .5 x 6 x 3 in . (140 x 150 x 70 mm)
5 .5 lbs . (2 .5 kg)
Source to skin distance
8 in . (200 mm) 12 in . optional (300 mm)
Generator performance
0 .5 kW
Nominal voltage (Accuracy)
50 to 70 kVp (±10%)
(Values below 60 kV are not in-
tended for human use in USA and Canada)
Nominal current (Accuracy)
4 to 7 mA (±20%)
X-ray tube cooling
Automatically controlled
≥ 122
°F (50 °C)
(Air cooling: optional)
Inherent filtration
1 .0 mm Al
Total filtration
Min . 2 .0 mm Al
Focal spot size according to IEC 60336
0 .4 mm
X-ray field collimation
∅
2 .4 in ./1 .2 x 1 .6 in
2
(
∅
60 mm/30 x 40 mm
2)
Optional 0 .8 x 1 .2 in
2
(20 x 30 mm
2
)
Exposure time (Accuracy)
0 .04 to 2 seconds (±5% or ±20 ms)
Anode material
Tungsten
Anode angle
12 .5 degrees
Duty cycle
1:60 or greater
General technical data
A6350-15
A6350-24
A6350-33
Arm length
17 .7 in . (450 mm)
23 .6 in . (600 mm)
35 .4 in . (900 mm)
Storage height
46 in . (1165 mm)
Weight
55 lbs (24 .9 kg)
56 lbs (25 .4 kg)
59 lbs (26 .7 kg)
Ambient conditions
during operation
during storage & transport
Temperature
50 to 95 °F (+10 to +35 °C)
14 to 140 °F (-10 to +60 °C)
Relative humidity
30 to 75%
10 to 75%
Air pressure
25 to 31 inHg (860 to 1060 hPa) 25 to 31 inHg (860 to 1060 hPa)