background image

Air Techniques, Inc

.

Page 4

2.7  Only use original parts

• Only Air Techniques parts or accessories and 

special accessories specifically approved by 
Air Techniques may be used . 

• Only use original working parts and spare 

parts .

Air Techniques cannot accept any liabili-
ty for damage caused by the use of ac-
cessories and special accessories not 
specifically approved by Air Techniques 
or not using original working parts and 
spare parts .

2.8  Transport

The original packaging offers the optimum pro-
tection for the appliance during transport .
If required, the original packaging for the unit 
can be ordered at Air Techniques .

Air Techniques cannot accept any liabili-
ty for damage caused during transport 
by the use of unsuitable packaging, this 
is also valid during the warranty term .

• Only transport the appliance in its original 

packaging whenever possible .

• Keep the packing materials out of the reach of 

children . 

• Attach the transport locking devices again .
• Do not expose the device to any strong 

shocks .

• Do not bump or pull the unit .

2.9  Disposal

The equipment contains - in some of its parts 
- solid and liquid substances which must be dis-
posed of at appropriate recycling centers con-
forming to all local, state and federal regulations .
In particular, the equipment contains the follow-
ing materials and/or components:

Tubehead:

Non-biodegradable plastic materials, metals, 
glass, dielectric oil, lead, tungsten .

Other parts:

Non-biodegradable plastics, metals, printed 
circuits, and electronic components .

Air Techniques is not responsible for dis-
posal of the apparatus or parts thereof 
and for the related expenses .

Important Information

2.10 Model identification plate

Appliance

The model identification plate is located on the 
rear side of the control unit housing .

REF Order number
SN

Serial number

UDI Unique device identification code

X-ray tube head

The model identification plate is located on the 
rear side of the X-ray unit .

REF Order number
SN

Serial number

Summary of Contents for A6350-15

Page 1: ...Installation Instructions P R O V E C T A H D Intraoral Dental X ray System ...

Page 2: ......

Page 3: ...4 6 Site preparation 10 4 7 Support load requirements 11 4 8 Electrical power requirements 11 4 9 Wiring length and gauge requirements 11 4 10 Remote switch configurations 12 5 Installation 14 5 1 Installation procedure summary 14 5 2 Control unit disassembly for base unit access 15 5 3 Base unit lock side setup 16 5 4 Base unit wall mounting options 17 5 5 Installing the horizontal arm 18 5 6 Ins...

Page 4: ...ce i e unplug and disconnect from mains 1 Documentation 1 1 Warnings and symbols Warnings The warnings in this document are there to point out possible injury to persons or damage to machinery The following warning symbols are used General warning symbol Warning dangerous electrical voltage Warning X rays The warnings are structured as follows SIGNAL WORD Description of type and source of danger P...

Page 5: ... EMC for medical products see section 12 Informa tion on EMC according to EN 60601 1 2 1 2 Notes on copyright All circuits processes names software and appliances quoted are protected under industrial property rights Any reprinting of the technical documentation in whole or in part is subject to prior approval of Air Techniques being given in writing 2 Safety This unit has been so designed and dev...

Page 6: ...als out of the reach of children Attach the transport locking devices again Do not expose the device to any strong shocks Do not bump or pull the unit 2 9 Disposal The equipment contains in some of its parts solid and liquid substances which must be dis posed of at appropriate recycling centers con forming to all local state and federal regulations In particular the equipment contains the follow i...

Page 7: ...ir Techniques Inc 3 Overview Product description 1 2 4 3 5 6 7 1 X ray tube head 2 Cone collimator 3 Main power switch 4 Control unit with control panel 5 Handheld exposure button 6 Horizontal Arm 7 Scissor Arm ...

Page 8: ... kW Nominal voltage Accuracy 50 to 70 kVp 10 Values below 60 kV are not in tended for human use in USA and Canada Nominal current Accuracy 4 to 7 mA 20 X ray tube cooling Automatically controlled 122 F 50 C Air cooling optional Inherent filtration 1 0 mm Al Total filtration Min 2 0 mm Al Focal spot size according to IEC 60336 0 4 mm X ray field collimation 2 4 in 1 2 x 1 6 in2 60 mm 30 x 40 mm2 Op...

Page 9: ...68 mm 31 inches 787 mm 36 5 inches 927 mm 71 75 inches 1823 mm 8 6 inches 220 mm 35 4 inches 900 mm PN A6350 33 68 5 inches 1740 mm 73 inches 1854 mm 42 inches 1068 mm 48 inches 1219 mm 83 3 inches 2118 mm 8 6 inches 220 mm C D 46 in 1165 mm 11 4 in 290 mm 180 ο 280 ο 380 ο 46 in 1165 mm A B See Horizontal Arm Length Below 3 2 Dimensions E F 21 in 524 mm DIMENTIONS A THRU E SINGLE POST WALL MOUNTI...

Page 10: ...Turned in H Head Turned out 17 7 inches 450 mm PN A6350 15 29 2 inches 742 5 mm 32 1 inches 814 7 mm 35 0 inches 890 mm 23 6 inches 600 mm PN A6350 24 35 1 inches 892 5 mm 37 inches 940 mm 41 inches 1040 mm 35 4 inches 900 mm PN A6350 33 48 inches 1219 2 mm 50 inches 1270 mm 53 inches 1346 mm 10 6 in 269 mm ...

Page 11: ... Hammer Crimp tool Lubricating grease Multimeter Impact drills 4 4 FDA Reporting Form Every Equipment Installation Service Mainte nance should be reported using the respective forms and checklists as annexed at the end of this manual Additionally it is mandatory to report every installation by filling the Form FDA 2579 May 2010 and submitting it to FDA Electronic Product Reports Radiological Healt...

Page 12: ... Core 1 X ray emitter lock 1 Flat head screw M3x6 2 Supporting arm cover 1 Grub screw M5x25 1 Long supporting arm cover plug 1 Short supporting arm cover plug 1 Grub screw M8x8 1 Bottom horizontal arm cover 1 Steel wall plugs only for concrete walls M8x30 5 Screw M5x30 2 Screw M8x20 4 Screw M5x15 with spring washer 2 Screw M3x8 6 Cable tie 3 Bolt M8x25L with spring flat washers 4 ea ...

Page 13: ...space for movement of the arms in the extended condition Make sure that the location allows the Provecta HD to be used with ease for all possible imaging procedures on the patient with respect to the patient chair location Site Environment Requirements Provecta HD is designed for indoor usage It should not be subjected to direct sunlight for any extended duration Mount the Provecta HD unit away fr...

Page 14: ...ing Length and Gauge Requirements Maximum length of wire and minimum gauge wire AWG from the power panel box to the Base Unit Line Voltage 100 to 240 10 Exposure Current 8 Amp Main Fuse Rating 10 Amp Provecta HD External Wiring Schematic Control Unit with Control Panel Building Power Panel Box Remote Control Panel Remote Wallplate Push Button Exposure Switch Note 2 Note 3 50 ft 15 m Max RJ11 Conne...

Page 15: ... Remote Keypad Console using Control Panel Remote Mounting Kit A6400 and ATI Remote Wall Plate Push Button Exposure Switch Kit A6320 Control Unit without Control Panel Remote Control Panel Remote Wallplate Pushbutton Exposure Switch Internal Wiring CAT 6 with RJ45 Connectors Power Source Control Panel or Outlet 3 Wire Power Connection 50 ft 15 m Max RJ11 Connectors Configuration No 2 1 ATI Remote ...

Page 16: ...nal Wiring CAT 6 with RJ45 Connectors Power Source Control Panel or Outlet Control Panel 3 Wire Power Connection 50 ft 15 m Max Control Unit without Control Panel Remote Wallplate Pushbutton Exposure Switch Control Unit with Control Panel Power Source Control Panel or Outlet 3 Wire Power Connection 3 Wire Remote Connection Remote Wallplate Pushbutton Exposure Switches Control Unit with Control Pan...

Page 17: ... 4 inch post installation using mounting plate P N A6389 16 Inch on Center Mount installation using mounting plate P N A6341 2 Control unit disassembly for base unit access 3 Base unit lock slide setup 4 Base unit with exposure keypad installation 5 Horizontal arm and scissor arm installation 6 Tubehead installation 7 Determine optional remote switch configuration and install as necessary See kit ...

Page 18: ... below and disassemble the control unit by performing the following steps 1 Slide the control panel slightly horizontal up 2 Disconnect the rear keypad connector and remove the control panel from the control unit 3 Loosen the 4 screws securing the control unit cover and carefully thread the keypad connector through the control unit cover while removing the cover 4 Proceed to the lock slide setup f...

Page 19: ...es right and left Refer to the illustrations below and set the lock slide for the desired arm control by placing the lock slide in the associated configuration as shown Repositioning the lock slide is as easy removing the two securing screws placing the lock slide in position and securing the slide with the two screws 1 8 0 ο 90 ο Right 9 0 ο L e f t Lock Slide Setting Typical Default Setting at 1...

Page 20: ...gle wood 4 x 4 post or equivalent drywall wall stud support using either the Single Post Wall Mounting 4 x 4 inch post Plate installation kit P N A6336 or the 16 Inch Center to Center Wall Mounting Plate installation kit A6330 mounting methods Refer to the procedures provided with the corresponding mounting plate installation kit to install the plate and mount the base unit Single Post Wall Mounti...

Page 21: ...wn in the figure 2 Vertical leveling control Loosen 6 point wrench bolt by using the wrench and insert grub screw M8x8 and adjusted vertical level 2 Insert the horizontal arm shaft into the base unit making sure not to pinch the 3 Wire Harness 3 Mount the horizontal arm brake with 2 screws M5x30 4 Connect the three horizontal arm cables A B and C as shown below Top of Base Unit Horizontal Arm Hori...

Page 22: ... of the head of the hexagon screw 1 2 Wire Harness Scissor Arm Shaft Horizontal Arm 1 2 Hexagon Screw Set Screw Grub Screw 5 Attach the scissor arm brake with two M5x15 screws with spring washer on the horizontal arm 6 Connect the ground cable to the PE conductor by using the truss head screw 7 Connect the cables of the scissor arm and the horizontal arm and screw on the cover with the two M3x6 fl...

Page 23: ...scissor arm can relax during assembly work Make sure that the scissor arm is folded during all assembly work 5 7 Installing the X ray Emitter Mount the X ray emitter on the installed scissor arm by performing the following procedure NOTE Apply grease lightly to the X ray emitter shaft as necessary 1 Insert the X ray emitter shaft into the scissor arm making sure not to pinch the associated cables ...

Page 24: ...e green with yellow stripe wire to the ground terminal Fixed Connection optional Power Terminal Block Mains Cable supplied Power Terminal Block 5 9 Connecting the Handheld Exposure Button Connection of the handheld exposure button to the PCB connector is made by simply inserting the as sociated wire connector into the mating connector on the bottom of the PCB Handheld Exposure Button RJ45 Connecto...

Page 25: ...le through the control unit cover and align the cover with the base unit 2 Install and tighten the 4 screws A B C D securing the control unit cover 3 Connect the rear keypad connector of the control panel 4 Align the panel spine with the control unit cover center groove and slide the control panel down in place B C A D Control Unit Cover Control Panel Keypad Connector Cable Connect ...

Page 26: ...re necessary follow the adjustment procedure below 1 Place the scissor arm 32 5 degree from horizontal arm as shown in the first image below When the angle is correct the tool will guide into the arm easily Note Generator scissor arm should be vertically 2 Insert the 350 mm long 5 mm allen key and turn clockwise to increase tension 3 Place the scissor arm straight as shown in the second image belo...

Page 27: ...s done by the third party before connecting to the door interlock 1 Prepare the door contact cable as shown below 2 Attach the door contact cable into the connec tor CN408 at the rear of the control panel If you try to connect the door contact cable through the control Unit to the control panel make sure that you pull out the cable from the right location as shown in the figure 3 Make sure that Do...

Page 28: ...ations The tests of constancy that must be carried out at regular intervals by the personnel are based on the results of the acceptance test 6 2 Electrical safety check Carry out an electrical safety check according to all national regulations e g patient conductivity conductivity of housing Document the results 6 3 Switch unit on The display will show the standard values for X ray exposures or th...

Page 29: ...confirm In menu option kV Option use to scroll and then use to confirm 50 70 kV are permissible in countries with no X ray limit regulations In menu option X ray Count use to scroll and then use to confirm In the menu option Expo S W Set use to make the respective selection and use to confirm Under menu option Version the current firmware version can be called up Use to acknowledge Under menu opti...

Page 30: ...ucts for which an RMA has been issued may be returned to the customer at the discretion of Air Techniques Any item returned under warranty will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have been defective Air Techniques Inc is not liable for indirect or consequential damages or loss of any nature in connection with this equipment Dealer l...

Page 31: ...ssex Street Sydney NSW 2000 Australia www airtechniques com Merchandise Surface Disinfectant Enzymatic Cleaner Hand Sanitizer and Lotion Waterline Cleaner Evacuation System Cleaner Imaging Accessories Chemistry Processor Accessories For over 50 years Air Techniques has been a leading innovator and manufac turer of dental products Our priority is ensuring complete satisfaction by man ufacturing rel...