-2-

SENSOR REPLACEMENT

STEP 4)

Remove sensor from sensor cup and remove leads . Take
the new sensor and remove the metal spring.  Reattach
leads to the proper colored terminals on the new sensor.
Install new sensor into sensor cup.

STEP 5)

Reassemble monitor and install back into system.
Connect all external connections.  Allow monitor
to stabilize 30 minutes to 1 hour and recalibrate

.

STEP 3)

Remove end plate to gain access
to the sensor cup from outside the
housing.

STEP 1)

Disconnect all external connections.
Remove CO monitor from the unit.

STEP 2)

Remove the four screws from the monitor’s
left end plate.

Note: Alarm location may vary.

Replacement sensors are shipped with a metal spring installed between the electrodes.  

Do not

 remove the clip

until the sensor is to be installed into the monitor.

Note:  Currently, there are two types of sensors being shipped.  Please check your sensor
lead connections and follow sensor connection instructions below.

Summary of Contents for CO-91 Series

Page 1: ...hesapeake Va 23320 Telephone 757 424 3967 Toll Free 1 800 866 8100 Fax No 757 424 5348 http www airsystems cc e mail sales airsystems cc SENSOR REPLACEMENT AND CALIBRATION PROCEDURE FOR CO 91 SERIES CARBON MONOXIDE MONITORS Manual No MON045 Rev 2 November 2003 Air Systems International Inc Registered to ISO 9001 Certificate No A5033 ...

Page 2: ...alibrate STEP 3 Remove end plate to gain access to the sensor cup from outside the housing STEP 1 Disconnect all external connections Remove CO monitor from the unit STEP 2 Remove the four screws from the monitor s left end plate Note Alarm location may vary Replacement sensors are shipped with a metal spring installed between the electrodes Do not remove the clip until the sensor is to be install...

Page 3: ... regulator fully by turning the knob at least two 2 turns counterclockwise 8 Allow digital readout to stabilize approximately 15 30 seconds 9 Adjust zero pot adjustment screw clockwise to increase counterclockwise to decrease until a 00 reading is obtained 10 Turn off the regulator and disconnect the tubing from the zero air regulator 1 Place the on off test switch to the on position 2 Allow 30 se...

Page 4: ...centration ppm as printed on the calibration gas cylinder 6 Turn regulator off and repeat zero adjustment procedure above Display should return to a 00 reading THE MONITOR IS NOW CALIBRATED AND SHOULD BE RECALIBRATED MONTHLY OR IF ACCURACY IS QUESTIONABLE CHECK LOCAL REQUIREMENTS AND RECALIBRATE AS REQUIRED 1 Install regulator to the CO calibration gas cylinder 2 Connect the plug to the monitor 3 ...

Reviews: