This guide does not replace the user manual
Intended for all the persons authorized to adjust settings
Quick guide
It is recommended to carry out this test after each
patient circuit and / or expiratory valve replacement.
Press
Autotests
Close
the patient circuit,
validate
to start the testing
sequence (<2 min).
After the beep and the message,
remove
the cap to
end the tests.
•
Autotests
Contact
Air Liquide Medical Systems
Parc de Haute Technologie
6, rue Georges Besse
92 182 Antony Cedex - France
Phone : +33 (0)1 40 96 66 00
Hotline : 0820 14 63 59
•
Expiratory valve MONNAL EVA
Extraction :
press the eject
button
Single use
valve
Autoclavable
valve
Cleaning protocol
(see user manual)
Assembly instructions
It is recommended to use a bacteriological filter on
the inspiratory output, to limit the risk of retro-
contamination (see user manual).
•
Bacteriological filter
Assembly :
•
Integrate
the silicone disks
into the membrane
(1)
•
Reposition
the membrane in
the valve body
(2)
•
Integrate
the flow sensor
(3)
at the valve body
(2)
(1)
(2)
(3)
(3)
www.device.airliquidehealthcare.com
Y
N
03
35
52
-
Ju
in
20
18
ver
sio
n
Note:
pressing the
padlock
locks the
touchscreen.
To unlock =
press
the padlock and
confirm
with the rotary knob.