Air Liquide MONNAL T60 Quick Manual Download Page 2

Press

the button on the left hand side of the

device.

Switch on the respirator

Start ventilation

Perform 

chest compressions :

-

5 beeps 

occur. 

-

Parameters are displayed in 

pink

CPV is now on!

fCC :

The frequency of chest compressions.

% CC :

percentage of time spent on continuious

chest compressions.

P-P :

efficiency indicator (indirect representation

of the pressures transmitted to the chest), to
monitor

the

regularity

of

the

compression

amplitude.

VTi :

monitoring the volume insufflated at each

cycle simultaneously with chest compressions.

CPR dedicated monitoring

In case of Return of Spontaneous
Circulation

Switch

to ROSC mode thanks to the button on the 

top right corner of the screen.
If you decide to reinitiate the cardiac compressions
(ex : recurrence of the CA), go back to CPV mode. 

EtCO

2

monitoring

CO2 : maximum instantaneous EtCO

2

value

measured between two ventilation cycles.
The CO2 trend : the green curve shows the
average, over the last 2 minutes, of the
maximum

values

​​of

instantaneous

CO

2

recorded between 2 ventilation cycles.

1-

Connect

the IRMA

™ probe to the etCO

2

connection socket.

2- Connect the probe to its patient adapter

(a). 

A click informs of the good installation of the probe.

3. The 

LED turns green, the IRMA ™ probe is ready 

for use 

(b).

CPV CO

(with CO

2

Irma probe)  

4. 

Connect

the IRMA ™ probe, equipped with its 

adapter, to the Y part of the patient circuit 

(c) 

and 

then to the endotracheal tube of the patient 

(d).

5. Position 

the IRMA ™ probe so that it is on the top 

of the interface 

(e).

Press CPV button

Connect

directly the ventilator to the mask

instead of the BAVU and/or to the intubation
sensor.

Verify

that the etCO

2

monitoring block displays

data.