![Air Liquide MONNAL T60 Quick Manual Download Page 2](http://html1.mh-extra.com/html/air-liquide/monnal-t60/monnal-t60_quick-manual_2875002002.webp)
Press
the button on the left hand side of the
device.
•
Switch on the respirator
•
Start ventilation
Perform
chest compressions :
-
5 beeps
occur.
-
Parameters are displayed in
pink
.
CPV is now on!
fCC :
The frequency of chest compressions.
% CC :
percentage of time spent on continuious
chest compressions.
P-P :
efficiency indicator (indirect representation
of the pressures transmitted to the chest), to
monitor
the
regularity
of
the
compression
amplitude.
VTi :
monitoring the volume insufflated at each
cycle simultaneously with chest compressions.
•
CPR dedicated monitoring
•
In case of Return of Spontaneous
Circulation
Switch
to ROSC mode thanks to the button on the
top right corner of the screen.
If you decide to reinitiate the cardiac compressions
(ex : recurrence of the CA), go back to CPV mode.
•
EtCO
2
monitoring
CO2 : maximum instantaneous EtCO
2
value
measured between two ventilation cycles.
The CO2 trend : the green curve shows the
average, over the last 2 minutes, of the
maximum
values
of
instantaneous
CO
2
recorded between 2 ventilation cycles.
1-
Connect
the IRMA
™ probe to the etCO
2
connection socket.
2- Connect the probe to its patient adapter
(a).
A click informs of the good installation of the probe.
3. The
LED turns green, the IRMA ™ probe is ready
for use
(b).
•
CPV CO
2
(with CO
2
Irma probe)
4.
Connect
the IRMA ™ probe, equipped with its
adapter, to the Y part of the patient circuit
(c)
and
then to the endotracheal tube of the patient
(d).
5. Position
the IRMA ™ probe so that it is on the top
of the interface
(e).
Press CPV button
Connect
directly the ventilator to the mask
instead of the BAVU and/or to the intubation
sensor.
Verify
that the etCO
2
monitoring block displays
data.