6
Power
ON
and
OFF
Pressing
any
key
(CLR,
MDE,
ENG,
ENT)
turns
unit
power
ON.
Pressing
CLR
&
ENT
simultaneously
turns
unit
power
OFF.
CLR
Pressing
CLR
clears
the
current
reading
or
zeros
the
transducer
if
no
reading
is
current.
MDE
Pressing
MDE
will
display
the
current
mode
for
a
1
second
period
of
time.The
unit
will revert
to
ACTIVE
DISPLAY
mode
after
that
time.
Depressing
the
MDE
key
a
second
time will
scroll
through
the
list
of
MODES
available:
PK,
AC 1PK, AC CAL TRACK
Depressing
any
other
key
will
return
the
unit
to
ACTIVE
DISPLAY
mode.
See
page
9
for more
descriptions.
ENG
Pressing
ENG
will
display
the
current
Unit
of
measure
for
a
1
second
period.
Pressing ENG
while
the
Units
of
Measure
are
displayed
will
scroll
through
the
available
units (Kgf
M,
Kgf
Cm,
gf
Cm,
Ncm,
Nm,
FT
LB,
IN
LB,
IN
OZ).
Pressing
any
other
key
will
return to
SAMPLE
mode.
ENT
Pressing
ENT
will
send
the
current
reading
to
the
serial
port
and
clears
the
screen.
MDE
&
ENG
Pressing
MDE
&
ENG
buttons
simultaneously
displays
the
Program
Menu.
See
Program Menu
section
for
details.
ACTIVE
DISPLAY
Active
Display
Mode
is
the
normal
operating
mode
used
by
the
Auditor
Torque
Cube™
for measurement.
When
a
measurement
is
taken,
the
transducer
reading
or
peak
is
displayed
for
the
length
of
time
set
by
Auto
Clear
(see
Program
Menu).
If
High
and
Low
limits
have
been
preset
(see
Program
Menu)
then
the
LED
light
will
show
Green
for
a sample
within
the
limits,
or
Red
for
one
that
is
outside
the
preset
limits.
In
Peak
and
1
st
Peak
modes,
the
LED
will
illuminate
Green
upon
completion
of
the
reading
if
no
limits are
preset
in
the
unit.
No
LED
illumination
will
occur
during
Track
mode.
BASIC OPERATION