9 - Remote Operation
41
QPX750 Instruction Manual
9.9.2
STATus Subsystem
·
:
STATus:OPERation:CONDition?
Query the Status Operation Condition register.
See Status reporting section for details. The content
is not deleted after being read as it indicates the
current operating status
·
:STATus:OPERation:ENABle <NRF>
Set Status Operation Enable Register to
<NRF>
9.9.3
MMEMory Subsystem
·
[MMEMory:]COPY:SETUP < Unquoted String1 >
Copy an instrument setup from a source of to a
destination The keyword LIVE represents the
current operating parameters of the instrument
and may be a source or destination. For example, to
copy the current operating parameters to a setup
called “mysetup” on the internal storage use:
:[MMEMory:]COPY:SETUP LIVE,mysetup Setup file
names have a maximum length of 8 characters.
9.9.4
SOURce Subsystem
·
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRF>
Set the output voltage to <NRF>
NV|UV|MV|KV|MAV
·
[SOURce:]VOLTage:VERify[:LEVel][:IMMediate][:AMPLitude] <NRF>
Set the output voltage to <NRF>
NV|UV|MV|V|KV|MAV with Verification
·
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRF>
Set the output current limit to <NRF>
NA|UA|MA|A
·
[SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] <NRF >
Set the power limit to <NRF>
NW|UW|MW|W|KW|MAW|GW
·
[SOURce:]VOLTage:RANGe <NR1>
Set the output range to <NRF> where <NRF>
= 1 for 0 - 50V Range and <NRF> = 2 for 0 - 80V
Range. Note: Trying to change Range to 0 - 50V
when Voltage of DC Output is set above 50V is
illegal and will generate execution error of 103.
·
[SOURce:]CURRent:PROTection[:LEVel] <NRF>
Set the Over Current Protection trip point to <NRF>
NA|UA|MA|A
·
[SOURce:]CURRent:PROTection:TRIPped?
Query the Over Current Protection status of the
unit. Response is of the format 0 if the OCP tripped
has not occurred and 1 if OCP tripped has occurred.