Page 42
Voltage Programming
Speed MX100QP only:
(Typical figures)
Maximum time required for output to settle within 1% of its total excursion
(for resistive load). Excludes command processing time.
90% Load
No Load
90% Load
No Load
35V 3A
16V 6A
70V 3A
35V 6A
Up
Up
Up
Up
10ms
10ms
25ms
10ms
10ms
10ms
12ms
10ms
Down
Down
Down
Down
60ms
10ms
80ms
20ms
550ms
350ms
850ms
550ms
Output Protection:
Output will withstand an applied forward voltage of up to 50V (O/P1, O/P2),
or 80V (O/P3, O/P4). Reverse protection by diode clamp for reverse
currents up to 3A.
Over-voltage Protection
(OVP) Trip:
Outputs 1&2: 1V to 40V. Output 3&4: 1V to 80V. Output trips off for OVP.
Resolution 100mV. Response time: typically 100us. Accuracy: ± (2% + 0.5V)
Over-current Protection
(OCP) Trip:
Measure-and-compare over-current protection is implemented in firmware.
Output trips off for OCP. Setting resolution: 10mA. Response time: typically
500ms. Accuracy: ± (0.3% + 2digits)
Over-temperature
Protection (OTP) Trip:
The output will be tripped off if a fault or blocked ventilation causes the
internal temperature to rise excessively.
METER SPECIFICATIONS (Each Output)
Voltage/Current Meters
5 digit meters for O/P 1 & O/P2; O/P 3 & O/P 4, 4 digit voltage meters
O/P 3 & O/P 4(70V).
Voltage
Resolution: O/P1, O/P2, O/P3 & O/P4 (35V): 1mV; O/P3 & O/P4 (70V):
10mV
Accuracy: 0.1% of reading ± 2 digits
Current
Resolution: 0.1mA
Accuracy: 0.3% of reading ± 5 digits
Current Meter Averaging User selectable On/Off per output with High, Medium or Low settings
V x A:
O/P1 & O/P2: 5-digits. Resolution 0.001W to 100W, 0.01W above 100W.
O/P3 and O/P4: 5-digits. Resolution 0.01W. Accuracy: 0.5% of reading ±
3 digits
SETTING MEMORY STORES
Stores for Individual
Outputs:
50 store positions for each output. Values stored are Range, Voltage,
Current, OVP and OCP
Stores for All Outputs:
50 store positions operating on all four outputs simultaneously.
Values stored are Range, Voltage, Current, OVP, OCP, Output On/Off,
Current Meter Averaging, Multi-On/Multi-Off Setup
USER INTERFACE
Display:
White on black backlit graphic LCD, 340 x 112 pixels.
Multiple font sizes and graphic icons.
Soft Keys:
Eight illuminated multi-function keys annotated from the display.
Home Screen:
Simultaneous display of meters and settings for all outputs.
Direct access to voltage or current setting for any output.
Individual Screens:
Display of meters and extended settings for an individual output (meters
in larger font). Direct access for all settings for that output.
Numeric Setting:
Floating point numeric entry of voltage, current, OVP or OCP.
Spin Wheel Setting:
Voltage, current and other parameters can be adjusted using the spin
wheel in quasi-analog fashion. Wheel can be disabled.
Menu Screen
System level functions are selected from a scrollable list.
Help Text
Multi-page help text is available for system level functions.