![Aim-TTI LD400 Instruction Manual Download Page 43](http://html.mh-extra.com/html/aim-tti/ld400/ld400_instruction-manual_2871871043.webp)
42
LD400 Status Model
*ESE
CME
OPC
QYE
PON
EXE
OR
OPERATION
COMPLETE
POWER
ON
COMMAND
ERROR
EXECUTION ERROR
NUMBER
QUERY ERROR
NUMBER
*ESR?
7
5
4
3
2
1
0
6
*PRE
&
&
&
&
&
7
5
4
3
2
1
0
6
ISE
SAT
LIM
OPW
FAULT
OR
&
&
&
&
&
&
FAULT CONDITION
RESPONSE MESSAGE
READY
600W WARNING
V BELOW DROPOUT
ISR?
OR
7
5
4
3
2
1
0
6
INST
SRQ
0
0
MAV
ESB
0
&
&
RSQ
MSS
*STB?
*SRE
INPUT SATURATED
DROP
INTR
0
0
OR
IST
= 0
/
EER?
QER?
= 0
/
7
5
4
3
2
1
0
6
7
5
4
3
2
1
0
6
ITE
OVP
TRIP
OCP
OR
&
&
&
&
FAULT TRIP
OVER CURRENT TRIP
OVER VOLTAGE TRIP
600W DURATION TRIP
ITR?
OPP
7
5
4
3
2
1
0
6
INPUT DISABLED
7
5
4
3
2
1
0
6
&
&
&
&
&
7
5
4
3
2
1
0
6
7
5
4
3
2
1
0
6
DIS
INPUT LIMITING
&
&
0
0
0
0
0
0
.
.
Register Summary
Set
Query
Name
†
ITR?
Input Trip Register
ITE
ITE?
Input Trip Enable Register
ISR?
Input State Register
ISE
ISE?
Input State Enable Register
†
EER?
Execution Error Register
†
QER?
Query Error Register
†
*ESR?
Standard Event Status Register
*ESE
*ESE?
Standard Event Status Enable Register
*STB?
Status Byte Register
*SRE
*SRE?
Status Byte Enable Register
*PRE
*PRE?
Parallel Poll Response Enable Register
†
These registers are cleared after being queried, or by the *CLS command.