10. Measuring menus
10.5 User measuring menu U1 data logger
User menu U1 can be freely configured by the
user using the AMR-Control software (see 10.6).
A data logger menu is provided as standard.
This menu can be used either on its own or just
like any measuring menu in conjunction with the
function menu
Data logger functions
(see 11.5).
The device status is displayed by certain sym-
bols in the status bar (see 8.2). Data acquisition
can be set to cyclic via the
cycle timer
. The available memory is displayed in
the function
memory capacity free
. This is omitted if neither internal memo-
ry nor a memory connector is available. The menu can then be used for output
via the interface to a printer or computer.
To start a cyclic measuring operation (if cycle >0) :
<START>
see 11.5.5
To initiate manual measured value scanning (if cycle =0) :
<MANU>
see 11.5.4
Example of a configured user measuring menu Bar chart
Or alternatively e.g. a user menu
Bar chart
could be configured using the AMR-Control soft-
ware (see 11.5). With the functions ´measured
value, small´ and ´bar chart´ 2 channels with
measured value and bar chart diagram can be
shown.
Measuring point selection
The 1st measuring channel is always the selected measuring point.
This can be selected directly as in any menu by means of
or
...
To change the other channels, the measuring point
PROG
and
must be selected as function by pressing key(s) :
or
...
The selected measuring point can now be changed :
< M ▲>
,
< M ▼>
...
To cancel the process of measuring point selection press :
<ESC>
To set the display range
the functions
analog start
and
analog end
in the
special functions
menu should be used; (see 12.12.3). Having selected these
functions they can be entered by pressing key(s)
PROG
and
... or they
can be entered directly on the appropriate axis; (see 8.5).
28 ALMEMO
®
2590
01: 21.67 °C Temperature
|¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹|¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹®®®¸®®®®®®®®®¸®®®®®®®®®¸®®®®®®®®®|
10.00 Ntc °C 30.0
11: 38.1 %H rel humidity
|¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹®®¸®®®®®®®®®|®®®®®®®®®®¸®®®®®®®®®¸®®®®®®®®®¸®®®®®®®®®¸®®®®®®®®|
0.0 % rH %H 100.0
MENU M ©©F FCT
C © REC COM l© ©l R01 * ´´´´´´µµµµ¶
02:
27.6
°C
NiCr Temperature º
Cycle timer
00:00:30 nS
Memory free
58.3 KB
START MENU M ©©F FCT