Dual and Dual Broadband NMR Probes
8
Dual and Dual Broadband NMR Probes Installation
01-999124-00 J 0606
Testing Probes on VnmrS and
UNITY
INOVA Systems
This section contains test and calibration procedures for the dual (
1
H/
13
C) and dual
broadband (
1
H–
19
F /
15
N–
31
P
)
probes on VnmrS and
UNITY
INOVA systems:
•
lists the tests appropriate for each probe type.
•
lists the test samples.
Table 2. Probe Test Order
Test
200 & 300
Dual
200, 300, 400,
Dual
Broadband
500
PFG
Dual
500 PFG
Dual
Broadband
*
*
*
*
H Observe 90° Pulse Width, page 9
*
*
*
*
H Spinning Resolution & Lineshape, page 10
*
*
*
*
*
*
*
*
PFG Profile and Recovery, page 11
PFG Probes
—
*
*
*
*
*
*
F Observe 90° Pulse Width, page 12
—
*
—
*
—
*
—
*
P Observe 90° Pulse Width, page 13
—
*
—
*
—
*
—
*
C Observe 90° Pulse Width, page 14
*
*
*
*
C Spinning Resolution & Lineshape, page 15
*
*
*
*
*
*
*
*
*
*
*
*
—
*
—
*
Table 3. Acceptance Tests and Samples
NMR Test
Sample
Part No.
1
H PW90,
1
H sensitivity
0.1% ethylbenzene, 0.01% TMS
99.89% deuterochloroform (CDCl3)
00-968120-70
1
H spinning res. & lineshape,
1
H spinning sidebands
1% chloroform in acetone–d
6
00-968120-89
PFG Profile and recovery
4-Hz 1% H
2
O, 99% D
2
O
01-901855-01
19
F PW90,
19
F sensitivity
0.05% trifluorotoluene in benzene–d
6
00-968120-82
31
P PW90,
31
P sensitivity
0.0485 M triphenylphosphate in CDCl
3
00-968120-87
13
C PW90
*
,
13
C sensitivity
*
40% p-dioxane in benzene–d
6
(ASTM) 0.2% Cr(acac)
(
00-968120-91
) can also be used.
40% p-dioxane in benzene–d
6
(ASTM)
00-968120-69
13
C spinning res. & lineshape,
13
C spinning sidebands
40% p-dioxane in benzene–d
6
(ASTM)
00-968120-69
15
N PW90
90% formamide in DMSO–d
6
00-968120-83