
Dual and Dual Broadband NMR Probes
16
Dual and Dual Broadband NMR Probes Installation
01-999124-00 J 0606
2.
Plot the spectrum again using a large enough value of
wp
to show all the spinning
sidebands, including fourth order.
3.
Measure spinning sideband amplitudes as a percentage of the peak.
Spinning sidebands occur at frequency intervals on either side of the central peak
equal to the spinning rate.
The standard test requires
nt=1
or
nt=4
. Do not repeat the test at
nt=16
if the
sideband specification is met at
nt=4
.
4.
Write the results for each probe in the forms provided in
13
C Sensitivity
The specification for the
13
C spinning resolution and lineshape test must be met before
performing this test.
1.
Enter
rtp('/vnmr/tests/C13sn')
su
.
2.
Tune the probe.
3.
Set the
13
C 90
°
pulse width to the value determined in
4.
Enter
ga
.
Wait at least 8 minutes between acquisitions to ensure that the C
6
D
6
lines are fully
relaxed.
5.
The signal of interest is the highest line of
the C
6
D
6
triplet (see
Use the cursors to select the best 1400 Hz
noise region between the dioxane and
benzene peaks.
6.
Enter
dsnmax(1400)
.
The computer calculates the S/N ratio.
7.
Answer
'y'
to plot the results.
8.
Write the results for each probe in the forms
provided in
.
15
N Observe 90° Pulse Width
Use this procedure to determine the
15
N PW90 value to be used in the
15
N sensitivity test.
1.
Enter
rtp('/vnmr/tests/N15sn') su
.
2.
Tune the probe observe channel without using the filter.
3.
Change
dpwr
to the
1
H
γ
H
2
value used in the procedure WALTZ
1
H Decoupling
Using High-Performance RF Waveform Generator. If a High-Performance RF
Waveform Generator is not present, use the procedure for WALTS
1
H Decoupling
Using Pre-programmed Pulse Modulation.
4.
Change
dmf
to 4*
γ
H
2
.
5.
Enter
ga
.
6.
Enter
dm='yyy' d1=45 su
.
7.
Determine the correct
tpwr
as follows:
a.
Enter
tpwr=50 pw=7 ga
.
Figure 6.
13
C Sensitivity
Hz
7000
8000
9000
10000
11000
12000
13000
14000
p-Dioxane
Benzene-d
6