26
User’s Guide U3025-90001
U3025AE10
Controlling the System with E8363/4C and N5230C
1. On the network analyzer select
Response
>
Cal Wizard
.
a. If using a mechanical cal kit, select
SmartCal (Guided Calibration)
>
Next
.
b. If using an ECal module, connect the ECal to an available PNA USB port and select
Use Electronic Calibration (ECal)
>
Next
.
2. Continue following the Cal Wizard prompts. In the Select Calibration Ports and ECal
Module dialog box press
Select All
,
or select the ports you are calibrating and
select
Next
.
Figure 26
12-Port Calibration
3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the
Cal Wizard prompts and press
Measure
after each connection.
4. At the Calibration Completed prompt select
Save As User Calset,
type the name and
Save
.
See
Figure 32 on page 34
.
5. After calibrating Test Set ports, use a quality load and short to verify the calibration on
each port or end of the test cable. Measure reflection and confirm the return loss is as
expected. If the result is not as expected, repeat the calibration without the Test Set and
ensure that the PNA is in standard (non-multiport) mode.
NOTE
If measurement errors occur, ensure the newest version of firmware is
installed on the PNA. Measurement errors can be a result of firmware
algorithms. Consult with Agilent Service or firmware web page for the latest
PNA Option 551 firmware revisions and history.
http:\\na.tm.agilent.compna\firmware\firmware.htm.
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html