Technical Information
RS 232/RS 485 Communication Description
17
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
197/224
4800 = 3
9600 = 4
(default = 4)
109
W
L
Pump life/ cycle time/
cycle number reset
To reset write ‘1’
110 R/W
L
Interlock
type
(default = 1)
Impulse = 0
Continuous = 1
111 R/W
L
Analog
output
type:
output voltage signal
pro-portional to
frequency or power
0 = frequency
1 = power
(default = 0)
120 R/W
N
Rotational
frequency
setting (Hz)
150 to 1100’
(default = 1100)
121 R/W
N
Maximum
rotational
frequency in Hz (active
only in Stop condition)
150 to 1100
(default = 1100)
122
R/W
L
Set vent valve on/off
(on = closed)
On = 1
Off = 0
(default = 1)
123
124
Reserved to Agilent service
125
R/W
L
Set the vent valve
operation
Automatic = 0 (see note 1.)
On command = 1
(see note 2.)
126
R/W
N
Vent valve opening
delay (expressed
in 0.2 sec)
0 to 65535 (corresponding
to 0 to 13107 sec)
130
Reserved to Agilent service
200
R
N
Pump current in mA dc
201
R
N
Pump voltage in Vdc
202
R
N
Pump power in W
(pump current x pump
voltage duty cycle)
203
R
N
Driving frequency in Hz
204
R
N
Pump temperature in
°C
205
R
N
Pump status
Stop = 0
Summary of Contents for TV 301-P