17
Technical Information
RS 232/RS 485 Communication Description
196/224
TV 301-P Navigator Controller User Manual
/ 87-901-001-01 (B)
Window-Meanings
Table 6
Window-Meanings
N. Read/Wr
ite
Dataty
pe
Description Admitted
Values
000
R/W
L
Start/Stop (in remote
mode the window is a
read only)
Start = 1
Stop = 0
008 R/W
L
Remote
(default)
or Serial configuration
Remote = 1
Serial = 0
(default = 1)
100
R/W
L
Soft Start (write only
in Stop condition)
YES = 1
NO = 0
101
R/W
N
Set Point type
0 = Frequency
1 = Current
2 = Time
(default = 0)
102
R/W
N
Set Point threshold
(expressed in Hz, mA
or s)
(default = 867)
103
R/W
N
Set Point delay: time
between the pump
start and the set point
check (seconds)
0 to 99999
(default = 0)
104 R/W
L
Set
Point
signal
activation type: the
signal can be "high
level active" or "low
level active"
0 = high level active
1 = low level active
(default = 0)
105
R/W
N
Set point hysteresis
(in % of threshold)
0 to 100
(default = 2)
106
R/W
L
Water cooling
0 = NO
1 = YES
107 R/W
L
Active
Stop
(write only in stop)
0 = NO
1 = YES
108
R/W
N
Baud rate
600 = 0
1200 = 1
2400 = 2